Semin Thromb Hemost 2024; 50(02): 253-270
DOI: 10.1055/s-0043-1772843
Review Article

Detecting Oxygenator Thrombosis in ECMO: A Review of Current Techniques and an Exploration of Future Directions

Jack Leerson
1   Department is Manufacturing, Materials and Mechatronics Engineering, School of Engineering, RMIT University, Melbourne, Victoria, Australia
2   Department of Manufacturing, CSIRO, Research Way, Clayton, Victoria, Australia
,
Andrew Tulloh
2   Department of Manufacturing, CSIRO, Research Way, Clayton, Victoria, Australia
,
Francisco Tovar Lopez
1   Department is Manufacturing, Materials and Mechatronics Engineering, School of Engineering, RMIT University, Melbourne, Victoria, Australia
,
Shaun Gregory
3   Department of Mechanical and Aerospace Engineering, Cardiorespiratory Engineering and Technology Laboratory, Monash University, Melbourne, Victoria, Australia
,
Hergen Buscher
4   Department of Intensive Care Medicine, St Vincent's Hospital, Sydney, Australia
,
Gary Rosengarten
1   Department is Manufacturing, Materials and Mechatronics Engineering, School of Engineering, RMIT University, Melbourne, Victoria, Australia
› Author Affiliations

Abstract

Extracorporeal membrane oxygenation (ECMO) is a life-support technique used to treat cardiac and pulmonary failure, including severe cases of COVID-19 (coronavirus disease 2019) involving acute respiratory distress syndrome. Blood clot formation in the circuit is one of the most common complications in ECMO, having potentially harmful and even fatal consequences. It is therefore essential to regularly monitor for clots within the circuit and take appropriate measures to prevent or treat them. A review of the various methods used by hospital units for detecting blood clots is presented. The benefits and limitations of each method are discussed, specifically concerning detecting blood clots in the oxygenator, as it is concluded that this is the most critical and challenging ECMO component to assess. We investigate the feasibility of solutions proposed in the surrounding literature and explore two areas that hold promise for future research: the analysis of small-scale pressure fluctuations in the circuit, and real-time imaging of the oxygenator. It is concluded that the current methods of detecting blood clots cannot reliably predict clot volume, and their inability to predict clot location puts patients at risk of thromboembolism. It is posited that a more in-depth analysis of pressure readings using machine learning could better provide this information, and that purpose-built imaging could allow for accurate, real-time clotting analysis in ECMO components.



Publication History

Article published online:
28 August 2023

© 2023. Thieme. All rights reserved.

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA

 
  • References

  • 1 Extracorporeal Life Support Organization (ELSO). ECLS International Summary of Statistics. Published October 19, 2022. Accessed February 16, 2023 at: https://www.elso.org/Registry/InternationalSummaryandReports/InternationalSummary.aspx
  • 2 Thiagarajan RR, Barbaro RP, Rycus PT. et al; ELSO member centers. Extracorporeal Life Support Organization Registry International Report 2016. ASAIO J 2017; 63 (01) 60-67
  • 3 Makdisi G, Wang IW. Extra corporeal membrane oxygenation (ECMO) review of a lifesaving technology. J Thorac Dis 2015; 7 (07) E166-E176
  • 4 Ramanathan K, Antognini D, Combes A. et al. Planning and provision of ECMO services for severe ARDS during the COVID-19 pandemic and other outbreaks of emerging infectious diseases. Lancet Respir Med 2020; 8 (05) 518-526
  • 5 Gajkowski EF, Herrera G, Hatton L, Velia Antonini M, Vercaemst L, Cooley E. ELSO guidelines for adult and pediatric extracorporeal membrane oxygenation circuits. ASAIO J 2022; 68 (02) 133-152
  • 6 Sy E, Sklar MC, Lequier L, Fan E, Kanji HD. Anticoagulation practices and the prevalence of major bleeding, thromboembolic events, and mortality in venoarterial extracorporeal membrane oxygenation: a systematic review and meta-analysis. J Crit Care 2017; 39: 87-96
  • 7 Nunez JI, Gosling AF, O'Gara B. et al. Bleeding and thrombotic events in adults supported with venovenous extracorporeal membrane oxygenation: an ELSO registry analysis. Intensive Care Med 2022; 48 (02) 213-224
  • 8 Zangrillo A, Landoni G, Biondi-Zoccai G. et al. A meta-analysis of complications and mortality of extracorporeal membrane oxygenation. Crit Care Resusc 2013; 15 (03) 172-178
  • 9 Pan KC, McKenzie DP, Pellegrino V, Murphy D, Butt W. The meaning of a high plasma free haemoglobin: retrospective review of the prevalence of haemolysis and circuit thrombosis in an adult ECMO centre over 5 years. Perfusion 2016; 31 (03) 223-231
  • 10 Lubnow M, Philipp A, Foltan M. et al. Technical complications during veno-venous extracorporeal membrane oxygenation and their relevance predicting a system-exchange–retrospective analysis of 265 cases. PLoS One 2014; 9 (12) e112316
  • 11 Kohs TCL, Liu P, Raghunathan V. et al. Severe thrombocytopenia in adults undergoing extracorporeal membrane oxygenation is predictive of thrombosis. Platelets 2022; 33 (04) 570-576
  • 12 McMichael ABV, Ryerson LM, Ratano D, Fan E, Faraoni D, Annich GM. 2021 ELSO adult and pediatric anticoagulation guidelines. ASAIO J 2022; 68 (03) 303-310
  • 13 Zhang M, Pauls JP, Bartnikowski N. et al. Anti-thrombogenic surface coatings for extracorporeal membrane oxygenation: a narrative review. ACS Biomater Sci Eng 2021; 7 (09) 4402-4419
  • 14 Bartlett R, Arachichilage DJ, Chitlur M. et al. The history of extracorporeal membrane oxygenation and the development of extracorporeal membrane oxygenation anticoagulation. Semin Thromb Hemost 2024; 50 (01) 81-90
  • 15 Figueroa Villalba CA, McMullan DM, Reed RC, Chandler WL. Thrombosis in extracorporeal membrane oxygenation (ECMO) circuits. ASAIO J 2022; 68 (08) 1083-1092
  • 16 Dalton HJ, Reeder R, Garcia-Filion P. et al; Eunice Kennedy Shriver National Institute of Child Health and Human Development Collaborative Pediatric Critical Care Research Network. Factors associated with bleeding and thrombosis in children receiving extracorporeal membrane oxygenation. Am J Respir Crit Care Med 2017; 196 (06) 762-771
  • 17 Murphy DA, Hockings LE, Andrews RK. et al. Extracorporeal membrane oxygenation-hemostatic complications. Transfus Med Rev 2015; 29 (02) 90-101
  • 18 Thomas J, Kostousov V, Teruya J. Bleeding and thrombotic complications in the use of extracorporeal membrane oxygenation. Semin Thromb Hemost 2018; 44 (01) 20-29
  • 19 Abruzzo A, Gorantla V, Thomas SE. Venous thromboembolic events in the setting of extracorporeal membrane oxygenation support in adults: a systematic review. Thromb Res 2022; 212: 58-71
  • 20 Sidebotham D, McGeorge A, McGuinness S, Edwards M, Willcox T, Beca J. Extracorporeal membrane oxygenation for treating severe cardiac and respiratory failure in adults: part 2-technical considerations. J Cardiothorac Vasc Anesth 2010; 24 (01) 164-172
  • 21 Kaesler A, Hesselmann F, Zander MO. et al. Technical indicators to evaluate the degree of large clot formation inside the membrane fiber bundle of an oxygenator in an in vitro setup. Artif Organs 2019; 43 (02) 159-166
  • 22 Lequier L, Horton SB, McMullan DM, Bartlett RH. Extracorporeal membrane oxygenation circuitry. Pediatr Crit Care Med 2013; 14 (5, Suppl 1): S7-S12
  • 23 Alfred Health. Haemolysis. ECMO guideline. Accessed May 29, 2022 at: https://ecmo.icu/daily-care-haemolysis/?parent=menuautoanchor-32&def=true
  • 24 Lehle K, Philipp A, Gleich O. et al. Efficiency in extracorporeal membrane oxygenation-cellular deposits on polymethylpentene membranes increase resistance to blood flow and reduce gas exchange capacity. ASAIO J 2008; 54 (06) 612-617
  • 25 Lehle K, Philipp A, Müller T. et al. Flow dynamics of different adult ECMO systems: a clinical evaluation. Artif Organs 2014; 38 (05) 391-398
  • 26 Sarathy S, Turek JW, Chu J, Badheka A, Nino MA, Raghavan ML. Flow monitoring of ECMO circuit for detecting oxygenator obstructions. Ann Biomed Eng 2021; 49 (12) 3636-3646
  • 27 Mossadegh C. Monitoring the ECMO. Nursing Care and ECMO 2017; 45-70
  • 28 Evans CF, Li T, Mishra V. et al. Externally visible thrombus partially predicts internal thrombus deposition in extracorporeal membrane oxygenators. Perfusion 2017; 32 (04) 301-305
  • 29 Alfred Health. ECMO circuit change. ECMO Guideline. Accessed June 02, 2022 at: https://ecmo.icu/procedures-ecmo-circuit-change/?parent=Daily
  • 30 Lubnow M, Philipp A, Dornia C. et al. D-dimers as an early marker for oxygenator exchange in extracorporeal membrane oxygenation. J Crit Care 2014; 29 (03) 473.e1-473.e5
  • 31 Panigada M, L'Acqua C, Passamonti SM. et al. Comparison between clinical indicators of transmembrane oxygenator thrombosis and multidetector computed tomographic analysis. J Crit Care 2015; 30 (02) 441.e7-441.e13
  • 32 Schaadt J. Oxygenator thrombosis: an international phenomenon. Perfusion 1999; 14 (06) 425-435
  • 33 Dornia C, Philipp A, Bauer S. et al. D-dimers are a predictor of clot volume inside membrane oxygenators during extracorporeal membrane oxygenation. Artif Organs 2015; 39 (09) 782-787
  • 34 Weitz JI, Fredenburgh JC, Eikelboom JW. A test in context: D-dimer. J Am Coll Cardiol 2017; 70 (19) 2411-2420
  • 35 Bates SM. D-dimer assays in diagnosis and management of thrombotic and bleeding disorders. Semin Thromb Hemost 2012; 38 (07) 673-682
  • 36 Wells PS. Integrated strategies for the diagnosis of venous thromboembolism. J Thromb Haemost 2007; 5 (Suppl. 01) 41-50
  • 37 Kamal AH, Tefferi A, Pruthi RK. How to interpret and pursue an abnormal prothrombin time, activated partial thromboplastin time, and bleeding time in adults. Mayo Clin Proc 2007; 82 (07) 864-873
  • 38 Horton S, Augustin S. Activated clotting time (ACT). In: Monagle P. ed. Haemostasis: Methods and Protocols. Methods in Molecular Biology. Totowa, NJ: Humana Press; 2013: 155-167
  • 39 Herrick S, Blanc-Brude O, Gray A, Laurent G. Fibrinogen. Int J Biochem Cell Biol 1999; 31 (07) 741-746
  • 40 Passmore MR, Fung YL, Simonova G. et al. Evidence of altered haemostasis in an ovine model of venovenous extracorporeal membrane oxygenation support. Crit Care 2017; 21 (01) 191
  • 41 Lowe GD, Rumley A, Mackie IJ. Plasma fibrinogen. Ann Clin Biochem 2004; 41 (Pt 6): 430-440
  • 42 Packham MA. Role of platelets in thrombosis and hemostasis. Can J Physiol Pharmacol 1994; 72 (03) 278-284
  • 43 Jurk K, Kehrel BE. Platelets: physiology and biochemistry. Semin Thromb Hemost 2005; 31 (04) 381-392
  • 44 Di Nardo M, Merli P, Cecchetti C, Pasotti E, Bertaina A, Locatelli F. Progressive increase in D-dimer levels during extracorporeal membrane oxygenation can predict membrane oxygenator failure in children given hematopoietic stem cell transplantation?. J Crit Care 2016; 31 (01) 262-263
  • 45 Gauer RL, Braun MM. Thrombocytopenia. Am Fam Physician 2012; 85 (06) 612-622
  • 46 Francis JL, Groce III JB, Group HC. Heparin Consensus Group. Challenges in variation and responsiveness of unfractionated heparin. Pharmacotherapy 2004; 24 (8, Pt 2): 108S-119S
  • 47 Irby K, Swearingen C, Byrnes J, Bryant J, Prodhan P, Fiser R. Unfractionated heparin activity measured by anti-factor Xa levels is associated with the need for extracorporeal membrane oxygenation circuit/membrane oxygenator change: a retrospective pediatric study. Pediatr Crit Care Med 2014; 15 (04) e175-e182
  • 48 Vandiver JW, Vondracek TG. Antifactor Xa levels versus activated partial thromboplastin time for monitoring unfractionated heparin. Pharmacotherapy 2012; 32 (06) 546-558
  • 49 Whiting D, DiNardo JA. TEG and ROTEM: technology and clinical applications. Am J Hematol 2014; 89 (02) 228-232
  • 50 Henderson N, Sullivan JE, Myers J. et al. Use of thromboelastography to predict thrombotic complications in pediatric and neonatal extracorporeal membranous oxygenation. J Extra Corpor Technol 2018; 50 (03) 149-154
  • 51 Colman E, Yin EB, Laine G. et al. Evaluation of a heparin monitoring protocol for extracorporeal membrane oxygenation and review of the literature. J Thorac Dis 2019; 11 (08) 3325-3335
  • 52 Panigada M, E Iapichino G, Brioni M. et al. Thromboelastography-based anticoagulation management during extracorporeal membrane oxygenation: a safety and feasibility pilot study. Ann Intensive Care 2018; 8 (01) 7
  • 53 Basken R, Cosgrove R, Malo J. et al. Predictors of oxygenator exchange in patients receiving extracorporeal membrane oxygenation. J Extra Corpor Technol 2019; 51 (02) 61-66
  • 54 Drop JG, Erdem Ö, Wildschut ED. et al. Use of rotational thromboelastometry to predict hemostatic complications in pediatric patients undergoing extracorporeal membrane oxygenation: a retrospective cohort study. Res Pract Thromb Haemost 2021; 5 (05) e12553
  • 55 Binder NB, Depasse F, Mueller J. et al. Clinical use of thrombin generation assays. J Thromb Haemost 2021; 19 (12) 2918-2929
  • 56 Chandler WL. Coagulation activation during extracorporeal membrane oxygenation (ECMO). Thromb Res 2022; 211: 154-160
  • 57 Cartwright B, Bruce HM, Kershaw G. et al. Hemostasis, coagulation and thrombin in venoarterial and venovenous extracorporeal membrane oxygenation: the HECTIC study. Sci Rep 2021; 11 (01) 7975
  • 58 Nossair F, Mahajerin A, Hoang J, Diaz D, Nugent D. Promising biomarkers for the prediction of catheter-related venous thromboembolism in hospitalized children: an exploratory study. Pediatr Blood Cancer 2019; 66 (10) e27870
  • 59 Chlebowski MM, Baltagi S, Carlson M, Levy JH, Spinella PC. Clinical controversies in anticoagulation monitoring and antithrombin supplementation for ECMO. Crit Care 2020; 24 (01) 19
  • 60 Krivitski N, Galyanov G, Cooper D. et al. In vitro and in vivo assessment of oxygenator blood volume for the prediction of clot formation in an ECMO circuit (theory and validation). Perfusion 2018; 33 (1_suppl): 51-56
  • 61 Badheka A, Stucker SE, Turek JW, Raghavan ML. Efficacy of flow monitoring during ECMO. ASAIO J 2017; 63 (04) 496-500
  • 62 Montalti A, Belliato M, Gelsomino S. et al. Continuous monitoring of membrane lung carbon dioxide removal during ECMO: experimental testing of a new volumetric capnometer. Perfusion 2019; 34 (07) 538-543
  • 63 Sakurai H, Fujiwara T, Ohuchi K. et al. Novel application of indocyanine green fluorescence imaging for real-time detection of thrombus in a membrane oxygenator. Artif Organs 2021; 45 (10) 1173-1182
  • 64 Reynolds JS, Troy TL, Mayer RH. et al. Imaging of spontaneous canine mammary tumors using fluorescent contrast agents. Photochem Photobiol 1999; 70 (01) 87-94
  • 65 Hope-Ross M, Yannuzzi LA, Gragoudas ES. et al. Adverse reactions due to indocyanine green. Ophthalmology 1994; 101 (03) 529-533
  • 66 Edwards Lifesciences. TruWave pressure transducers | Edwards Lifesciences. Accessed November 22, 2022 at: https://www.edwards.com/gb/devices/Pressure-Monitoring/Transducer
  • 67 Shashikumar SP, Stanley MD, Sadiq I. et al. Early sepsis detection in critical care patients using multiscale blood pressure and heart rate dynamics. J Electrocardiol 2017; 50 (06) 739-743
  • 68 Mayaud L, Tarassenko L, Annane D, Clifford G. Predictive power of heart rate complexity to estimate severity in severe sepsis patients. J Crit Care 2013; 28 (06) e37
  • 69 Nemati S, Holder A, Razmi F, Stanley MD, Clifford GD, Buchman TG. An interpretable machine learning model for accurate prediction of sepsis in the ICU. Crit Care Med 2018; 46 (04) 547-553
  • 70 Hatib F, Jian Z, Buddi S. et al. Machine-learning algorithm to predict hypotension based on high-fidelity arterial pressure waveform analysis. Anesthesiology 2018; 129 (04) 663-674
  • 71 La Rovere MT, Pinna GD, Raczak G. Baroreflex sensitivity: measurement and clinical implications. Ann Noninvasive Electrocardiol 2008; 13 (02) 191-207
  • 72 Wu K, Feng Y, Xu Y. Research on method for detecting pipeline blockages based on fluid oscillation theory. Energies 2022; 15 (15) 5373
  • 73 Zouari F, Blåsten E, Louati M, Ghidaoui MS. Internal pipe area reconstruction as a tool for blockage detection. Journal of Hydraulic Engineering. 2019;145(06):
  • 74 Sondhi MM, Gopinath B. Determination of vocal-tract shape from impulse response at the lips. J Acoust Soc Am 1971; 49 (06) 1867-1873
  • 75 Fuchs G, Berg N, Eriksson A, Prahl Wittberg L. Detection of thrombosis in the extracorporeal membrane oxygenation circuit by infrasound: proof of concept. Artif Organs 2017; 41 (06) 573-579
  • 76 Kaufmann F, Hörmandinger C, Stepanenko A. et al. Acoustic spectral analysis for determining pump thrombosis in rotary blood pumps. ASAIO J 2014; 60 (05) 502-507
  • 77 Hubbert L, Sundbom P, Loebe M, Peterzén B, Granfeldt H, Ahn H. Acoustic analysis of a mechanical circulatory support. Artif Organs 2014; 38 (07) 593-598
  • 78 Dornia C, Philipp A, Bauer S. et al. Analysis of thrombotic deposits in extracorporeal membrane oxygenators by multidetector computed tomography. ASAIO J 2014; 60 (06) 652-656
  • 79 Conway RG, Zhang J, Jeudy J. et al. Computed tomography angiography as an adjunct to computational fluid dynamics for prediction of oxygenator thrombus formation. Perfusion 2021; 36 (03) 285-292
  • 80 Dornia C, Philipp A, Bauer S. et al. Visualization of thrombotic deposits in extracorporeal membrane oxygenation devices using multidetector computed tomography: a feasibility study. ASAIO J 2013; 59 (04) 439-441
  • 81 D'Onofrio C, van Loon R, Rolland S. et al. Three-dimensional computational model of a blood oxygenator reconstructed from micro-CT scans. Med Eng Phys 2017; 47: 190-197
  • 82 Bultreys T, Boone MA, Boone MN. et al. Fast laboratory-based micro-computed tomography for pore-scale research: illustrative experiments and perspectives on the future. Adv Water Resour 2016; 95: 341-351
  • 83 Dierick M, Van Loo D, Masschaele B. et al. Recent micro-CT scanner developments at UGCT. Nucl Instrum Methods Phys Res B 2014; 324: 35-40
  • 84 Clark DP, Badea CT. Advances in micro-CT imaging of small animals. Phys Med 2021; 88: 175-192
  • 85 Staaf E, Brehwens K, Haghdoost S. et al. Micronuclei in human peripheral blood lymphocytes exposed to mixed beams of X-rays and alpha particles. Radiat Environ Biophys 2012; 51 (03) 283-293
  • 86 Kaesler A, Rudawski FL, Zander MO. et al. In-vitro visualization of thrombus growth in artificial lungs using real-time X-ray imaging: a feasibility study. Cardiovasc Eng Technol 2022; 13 (02) 318-330
  • 87 Sakota D, Murashige T, Kosaka R, Fujiwara T, Nishida M, Maruyama O. Real-time observation of thrombus growth process in an impeller of a hydrodynamically levitated centrifugal blood pump by near-infrared hyperspectral imaging. Artif Organs 2015; 39 (08) 714-719
  • 88 Sakota D, Murashige T, Kosaka R. et al. Noninvasive optical imaging of thrombus formation in mechanical circulatory support devices. Journal of Biorheology. 2016; 30 (01) 6-12
  • 89 Sakota D, Kosaka R, Nishida M, Maruyama O. Optical aggregometry of red blood cells associated with the blood-clotting reaction in extracorporeal circulation support. J Artif Organs 2016; 19 (03) 241-248
  • 90 Sakota D, Fujiwara T, Ouchi K, Kuwana K, Yamazaki H, Maruyama O. Development of an optical detector of thrombus formation on the pivot bearing of a rotary blood pump. Artif Organs 2016; 40 (09) 834-841
  • 91 Sakota D, Fujiwara T, Ohuchi K. et al. Development of a real-time and quantitative thrombus sensor for an extracorporeal centrifugal blood pump by near-infrared light. Biomed Opt Express 2017; 9 (01) 190-201
  • 92 Fujiwara T, Sakota D, Ohuchi K. et al. Optical dynamic analysis of thrombus inside a centrifugal blood pump during extracorporeal mechanical circulatory support in a porcine model. Artif Organs 2017; 41 (10) 893-903
  • 93 Morita N, Sakota D, Oota-Ishigaki A. et al. Real-time, non-invasive thrombus detection in an extracorporeal circuit using micro-optical thrombus sensors. Int J Artif Organs 2021; 44 (08) 565-573
  • 94 Matsuhashi Y, Sameshima K, Yamamoto Y, Umezu M, Iwasaki K. Real-time visualization of thrombus formation at the interface between connectors and tubes in medical devices by using optical coherence tomography. PLoS One 2017; 12 (12) e0188729
  • 95 Lensing AWA, Prandoni P, Brandjes D. et al. Detection of deep-vein thrombosis by real-time B-mode ultrasonography. N Engl J Med 1989; 320 (06) 342-345
  • 96 Shiloh AL, McPhee C, Eisen L, Koenig S, Millington SJ. Better with ultrasound: detection of DVT. Chest 2020; 158 (03) 1122-1127
  • 97 Ierardi AM, Coppola A, Fusco S. et al. Early detection of deep vein thrombosis in patients with coronavirus disease 2019: who to screen and who not to with Doppler ultrasound?. J Ultrasound 2021; 24 (02) 165-173
  • 98 Reddick M, Kalva S. Non-invasive imaging of dialysis access circuit. In: Wu S, Kalva S, Park H, Tan CS, Beathard GA. eds. Dialysis Access Management. New York, NY: Springer International Publishing; 2021: 95-114
  • 99 Siqueira MHS, Gatts CEN, da Silva RR, Rebello JMA. The use of ultrasonic guided waves and wavelets analysis in pipe inspection. Ultrasonics 2004; 41 (10) 785-797
  • 100 Li X, Liu Y, Liu Z. et al. A hydrate blockage detection apparatus for gas pipeline using ultrasonic focused transducer and its application on a flow loop. Energy Sci Eng 2020; 8 (05) 1770-1780
  • 101 Piao C, Kim SH, Lee JK, Choi WG, Kim YY. Non-invasive ultrasonic inspection of sludge accumulation in a pipe. Ultrasonics 2022; 119: 106602
  • 102 Attia ABE, Balasundaram G, Moothanchery M. et al. A review of clinical photoacoustic imaging: current and future trends. Photoacoustics 2019; 16: 100144
  • 103 Beard P. Biomedical photoacoustic imaging. Interface Focus 2011; 1 (04) 602-631
  • 104 Das D, Pramanik M. Combined ultrasound and photoacoustic imaging of blood clot during microbubble-assisted sonothrombolysis. J Biomed Opt 2019; 24 (12) 1-8
  • 105 Karpiouk AB, Aglyamov SR, Mallidi S. et al. Combined ultrasound and photoacoustic imaging to detect and stage deep vein thrombosis: phantom and ex vivo studies. J Biomed Opt 2008; 13 (05) 054061
  • 106 Das D, Sivasubramanian K, Rajendran P, Pramanik M. Label-free high frame rate imaging of circulating blood clots using a dual modal ultrasound and photoacoustic system. J Biophotonics 2021; 14 (03) e202000371
  • 107 Wabe T, Suzuki R, Fujii A, Uchida Y, Maruyama K, Uchida Y. Basic research on blood coagulation measurement of extracorporeal circulation circuit using photoacoustic imaging of LED light source. Sensors & Transducers. 2021; 253 (06) 1-8
  • 108 Ke H, Erpelding TN, Jankovic L, Liu C, Wang LV. Performance characterization of an integrated ultrasound, photoacoustic, and thermoacoustic imaging system. J Biomed Opt 2012; 17 (05) 056010
  • 109 Tang Y, Wu H, Klippel P. et al. Deep thrombosis characterization using photoacoustic imaging with intravascular light delivery. Biomed Eng Lett 2022; 12 (02) 135-145
  • 110 Li J, Kikuchi D, Sapkota A, Takei M. Quantitative evaluation of electrical parameters influenced by blood flow rate with multiple-frequency measurement based on modified Hanai mixture formula. Sens Actuators B Chem 2018; 268: 7-14
  • 111 Li J, Sapkota A, Kikuchi D, Sakota D, Maruyama O, Takei M. Red blood cells aggregability measurement of coagulating blood in extracorporeal circulation system with multiple-frequency electrical impedance spectroscopy. Biosens Bioelectron 2018; 112: 79-85
  • 112 Li J, Wan N, Wen J, Cheng G, He L, Cheng L. Quantitative detection and evaluation of thrombus formation based on electrical impedance spectroscopy. Biosens Bioelectron 2019; 141: 111437
  • 113 Xu Z, Yao J, Wang Z. et al. Development of a portable electrical impedance tomography system for biomedical applications. IEEE Sens J 2018; 18 (19) 8117-8124
  • 114 Sapkota A, Fuse T, Seki M, Maruyama O, Sugawara M, Takei M. Application of electrical resistance tomography for thrombus visualization in blood. Flow Meas Instrum 2015; 46: 334-340
  • 115 Chen H, Yao J, Yang L. et al. Development of a portable electrical impedance tomography device for online thrombus detection in extracorporeal-circulation equipment. IEEE Sens J 2021; 21 (03) 3653-3659
  • 116 Hoshino K, Muranishi K, Kawano Y. et al. Soluble fibrin is a useful marker for predicting extracorporeal membrane oxygenation circuit exchange because of circuit clots. J Artif Organs 2018; 21 (02) 196-200
  • 117 Umei N, Ichiba S, Genda Y, Mase H, Sakamoto A. Early predictors of oxygenator exchange during veno-venous extracorporeal membrane oxygenation: a retrospective analysis. Int J Artif Organs 2022; 45 (11) 927-935