Am J Perinatol 2011; 28(8): 659-666
DOI: 10.1055/s-0031-1276741
© Thieme Medical Publishers

Intraluminal Pressure in a Uterine Tamponade Balloon Is Curvilinearly Related to the Volume of Fluid Infused

Michael A. Belfort1 , Gary A. Dildy2 , Jose Garrido3 , G. Lance White1
  • 1Department of Obstetrics and Gynecology, Baylor College of Medicine, Houston, Texas
  • 2Maternal–Fetal Services of Utah, St. Mark's Hospital, Salt Lake City, Utah
  • 3Hospital de la Mujer Dominicana, Santa Domingo, Republica Dominicana
Further Information

Publication History

Publication Date:
15 April 2011 (online)

ABSTRACT

We studied the effect of incremental infusion of fluid volume in a tamponade balloon on intraluminal pressure and uterine blood flow. Following placental delivery, a tamponade balloon was inserted into the uterus and incrementally inflated. Intraluminal pressure was measured at incremental volumes. Ultrasound was used to determine positioning of the catheter, uterine wall thickness, and uterine artery velocity waveforms in eight patients. Pressure-volume relationship was estimated by regression analysis. Significance was p < 0.05. There was a significant exponential curvilinear relationship between balloon pressure and infused volume at the maximum volume for each subject (R = 0.64, p = 0.01). Doppler ultrasound showed that at or above 1000 mL inflation volume, 5/6 patients (83%) showed reversal of uterine artery diastolic flow. At maximal inflation volume, all of the patients with reversed diastolic flow had intraluminal pressure less than systolic blood pressure. Intraluminal pressure increases curvilinearly as volume of an intrauterine tamponade balloon is increased. The mechanism of action of tamponade balloons is likely related to a reduction in uterine artery perfusion pressure. Whether this is the result of direct compression of the artery in the lower segment or due to wall conformational changes is not clear.

REFERENCES

  • 1 Georgiou C. Balloon tamponade in the management of postpartum haemorrhage: a review.  BJOG. 2009;  116 748-757
  • 2 Cho Y, Rizvi C, Uppal T, Condous G. Ultrasonographic visualization of balloon placement for uterine tamponade in massive primary postpartum hemorrhage.  Ultrasound Obstet Gynecol. 2008;  32 711-713
  • 3 Georgiou C. Intraluminal pressure readings during the establishment of a positive “tamponade test” in the management of postpartum haemorrhage.  BJOG. 2010;  117 295-303
  • 4 Bieniarz J, Branda L A, Maqueda E, Morozovsky J, Caldeyro-Barcia R. Aortocaval compression by the uterus in late pregnancy. 3. Unreliability of the sphygmomanometric method in estimating uterine artery pressure.  Am J Obstet Gynecol. 1968;  102 1106-1115
  • 5 Cook Medical Website .Available at: https://ifu.cookmedical.com/ifuPub/searchIfu.jsf (accessed September 1, 2010)

Michael A BelfortM.B.B.Ch. M.D. Ph.D. 

Chairman and Professor, Department of Obstetrics and Gynecology, Baylor College of Medicine

1709 Dryden, Suite 1100, Houston, TX 77030

Email: belfort@bcm.edu

    >