Endoscopic ultrasound-guided antegrade stent deployment (EUS-AS) has recently been
performed for patients with inaccessible papilla, for example due to malignant duodenal
obstruction or surgically altered anatomy [1]
[2]
[3]. During EUS-AS, use of a fine-gauge delivery system may prevent bile leakage because
tract dilation may not be needed [4]. Additionally, to prevent stent misplacement or dislocation, a laser-cut-type uncovered
self-expandable metal stent (UCSEMS) may be favorable compared with a fully covered
metal stent (FCSEMS), although stent patency is shorter. Furthermore, cell width is
high, so stent patency may be insufficient. Recently, a novel laser-cut-type UCSEMS
(YABUSAME; Kaneka, Tokyo, Japan) has become available. Cell width is smaller than
in the conventional laser-cut-type UCSEMS in the part where bending force is applied
([Fig. 1]). In addition, the stent delivery system is only 5.4 Fr ([Fig. 2]). Transluminal stent insertion can thus be performed without additional tract dilation.
We describe herein technical tips for EUS-AS using this novel stent.
Fig. 1 Cell width of the novel uncovered self-expandable metal stent (YABUSAME; Kaneka,
Tokyo, Japan) is small compared with the conventional type, particularly in the part
where bending force is applied. a YABUSAME stent. b Conventional uncovered self-expandable metal stent.
Fig. 2 The stent delivery system is 5.4 Fr, with an extremely tapered tip.
First, the intrahepatic bile duct is punctured using a 19-G needle, and contrast medium
is injected. A 0.025-inch guidewire is then inserted into the common bile duct. Next,
an endoscopic retrograde cholangiopancreatography (ERCP) catheter is inserted into
the biliary tract, and contrast medium is injected to evaluate the stricture site.
After the guidewire is advanced into the intestine across the stricture site ([Fig. 3]), insertion of the novel stent delivery system is successfully performed without
additional tract dilation ([Fig. 4, ]
[Video 1]). Stent deployment from the lower bile duct to the upper bile duct is also successfully
performed ([Fig. 5]). Finally, metal stent deployment is performed from the intrahepatic bile duct to
the stomach.
Fig. 3 A 0.025-inch guidewire is successfully inserted into the intestine across the stricture
site.
Fig. 4 The stent delivery system is successfully inserted into the biliary tract without
additional tract dilation.
Video 1 Guidewire passage across the stricture site is successfully performed. Antegrade
stent deployment using a novel uncovered self-expandable metal stent is performed
without additional fistula dilation.
Fig. 5 Antegrade stent deployment is successfully performed.
This novel UCSEMS may be useful as an antegrade stent because of the favorable stent
patency and fine-gauge stent delivery system, although a prospective comparative study
is needed.
Endoscopy_UCTN_Code_TTT_1AS_2AD
Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques
in gastroenterological endoscopy. All papers include a high quality video and all
contributions are freely accessible online. Processing charges apply (currently EUR
375), discounts and wavers acc. to HINARI are available.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos