Am J Perinatol 2017; 34(10): 1026-1031
DOI: 10.1055/s-0037-1601440
Original Article
Thieme Medical Publishers, Inc. 333 Seventh Avenue, New York, NY 10001, USA

Temporal Artery Temperature Measurement in the Neonate

Mashette E. Syrkin-Nikolau
1   Department of Pediatrics, University of Iowa, Iowa City, Iowa
,
Karen J. Johnson
1   Department of Pediatrics, University of Iowa, Iowa City, Iowa
,
Tarah T. Colaizy
1   Department of Pediatrics, University of Iowa, Iowa City, Iowa
,
Ruthann Schrock
1   Department of Pediatrics, University of Iowa, Iowa City, Iowa
,
Edward F. Bell
1   Department of Pediatrics, University of Iowa, Iowa City, Iowa
› Author Affiliations
Further Information

Publication History

01 November 2016

28 February 2017

Publication Date:
10 April 2017 (online)

Abstract

Objective We compared an infrared temporal artery thermometer with our clinical standard axillary thermometer for temperature measurements in neonatal patients.

Study Design We measured temporal artery (Tta), axillary (Tax, clinical standard), and rectal (Tr, gold standard) temperatures of 49 infants. The difference between Tr and Tta was compared with that between Tr and Tax, and the data were analyzed based on bed type and postmenstrual age.

Results The mean Tta, Tax, and Tr were 37.16 (SD 0.36) °C, 36.61 (SD 0.30) °C, and 36.82 (SD 0.30) °C, respectively. The measurements by these methods were all significantly different. The mean Tr-Tax was 0.21 (SD 0.26) °C, and the mean Tr-Tta was −0.34 (SD 0.37) °C, indicating that Tax was closer to Tr than was Tta (p < 0.0001). Tta agreed more closely with Tr for infants in cribs than for those in incubators. Adjusting for bed type and body weight, with each week of postmenstrual age, the discrepancy between Tr-Tta and Tr-Tax decreased by 0.005°C (p = 0.034).

Conclusion Compared with the gold standard, Tr, Tta is not more accurate than Tax. The temporal artery thermometer was less accurate for infants in incubators than for infants in cribs. The accuracy of temporal artery temperature increased with postmenstrual age.

Note

The authors thank the patients who participated and their parents, who granted consent for their participation. The authors thank Frank H. Morriss, Jr., MD, for advice about the Bland-Altman analysis. Ms. Syrkin-Nikolau was supported by a summer research fellowship from the University of Iowa Carver College of Medicine, which was funded by NIH grant T35 HL007485. This work was also supported by grant UL1 RR024979, awarded from the NIH to support the University of Iowa Institute for Clinical and Translational Science. The temporal artery thermometer was purchased by the investigators and not provided by the manufacturer, who had no knowledge of the study or its results.


 
  • References

  • 1 Segnitz RH. Accidental transanal perforation of the rectum: an obscure cause of peritonitis in infancy and childhood. AMA J Dis Child 1957; 93 (03) 255-258
  • 2 Fonkalsrud EW, Clatworthy Jr HW. Accidental perforation of the colon and rectum in newborn infants. N Engl J Med 1965; 272: 1097-1100
  • 3 Frank JD, Brown S. Thermometers and rectal perforations in the neonate. Arch Dis Child 1978; 53 (10) 824-825
  • 4 Lau JT, Ong GB. Broken and retained rectal thermometers in infants and young children. Aust Paediatr J 1981; 17 (02) 93-94
  • 5 Mayfield SR, Bhatia J, Nakamura KT, Rios GR, Bell EF. Temperature measurement in term and preterm neonates. J Pediatr 1984; 104 (02) 271-275
  • 6 Exergen Corporation. Model TAT-5000 Reference Manual. p/n 818528 r5. Watertown, MA: Exergen Corporation, undated. http://www.exergen.com/medical/PDFs/TAT5000Manualr5-5rev-ecl.pdf . Accessed March 21, 2017
  • 7 Abrams RM, Royston JP, Humphrey SJ, Wolff HS. Thermal features of the female axilla. J Appl Physiol 1980; 49 (06) 1042-1046
  • 8 Karlberg P. The significance of depth of insertion of the thermometer for recording rectal temperatures. Acta Paediatr 1949; 38: 359-366
  • 9 Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1 (8476): 307-310
  • 10 Giavarina D. Understanding Bland Altman analysis. Biochem Med (Zagreb) 2015; 25 (02) 141-151
  • 11 Mayfield SR, Nakamura KT, Bhatia J, Rios GR, Bell EF. Tympanic membrane temperature of term and preterm neonates. Early Hum Dev 1984; 9 (03) 241-247
  • 12 Johnson KJ, Bhatia P, Bell EF. Infrared thermometry of newborn infants. Pediatrics 1991; 87 (01) 34-38
  • 13 Weiss ME. Tympanic infrared thermometry for fullterm and preterm neonates. Clin Pediatr (Phila) 1991; 30 (4, Suppl): 42-45 , discussion 49
  • 14 Weiss ME, Poeltler D, Gocka I. Infrared tympanic thermometry for neonatal temperature assessment. J Obstet Gynecol Neonatal Nurs 1994; 23 (09) 798-804
  • 15 Craig JV, Lancaster GA, Taylor S, Williamson PR, Smyth RL. Infrared ear thermometry compared with rectal thermometry in children: a systematic review. Lancet 2002; 360 (9333): 603-609
  • 16 Duran R, Vatansever Ü, Acunaş B, Süt N. Comparison of temporal artery, mid-forehead skin and axillary temperature recordings in preterm infants <1500 g of birthweight. J Paediatr Child Health 2009; 45 (7-8): 444-447
  • 17 Lee G, Flannery-Bergey D, Randall-Rollins K. , et al. Accuracy of temporal artery thermometry in neonatal intensive care infants. Adv Neonatal Care 2011; 11 (01) 62-70
  • 18 Haddad L, Smith S, Phillips KD, Heidel RE. Comparison of temporal artery and axillary temperatures in healthy newborns. J Obstet Gynecol Neonatal Nurs 2012; 41 (03) 383-388
  • 19 Jarvis M, Guy KJ, König K. Accuracy of infrared thermometers in very low birth weight infants and impact on newborn behavioural states. J Paediatr Child Health 2013; 49 (06) 471-474
  • 20 Sim MA, Leow SY, Hao Y, Yeo CL. A practical comparison of temporal artery thermometry and axillary thermometry in neonates under different environments. J Paediatr Child Health 2016; 52 (04) 391-396