CC BY-NC-ND 4.0 · Am J Perinatol
DOI: 10.1055/s-0043-1775562
Review Article

A Systematic Review and Meta-analysis of Two Negative Pressure Wound Therapy Devices to Manage Cesarean Section Incisions

Theodore Goldman
1   Obstetrics and Gynecology, Northwell Health, Huntington, New York
,
Ben Costa
2   Global Clinical and Medical Affairs, Smith + Nephew, Hull, United Kingdom
› Author Affiliations
Funding This study was funded by Smith & Nephew, Inc.
 

Abstract

This paper aims to evaluate whether there is a device-dependent effect on the reduction of surgical site complications in obese patients (body mass index [BMI] ≥ 30 kg/m2) undergoing cesarean section (C-section). PubMed, Embase, Cochrane Library, and ClinicalTrials.gov were searched for the period, January 2011 to September 2021. English language articles describing a randomized controlled trial (RCT) that compared either a −80 or −125 mm Hg single-use negative pressure wound therapy (sNPWT) device to standard dressings in obese (BMI ≥ 30 kg/m2) patients undergoing C-section were included. Conference abstracts and “terminated” RCTs with published results were deemed eligible for inclusion. The primary outcome of interest was surgical site infection (SSI), classified as composite, superficial, or deep. Secondary outcomes assessed included seroma, dehiscence, hematoma, bleeding, reoperation, readmission, blistering, and (composite) wound complications. A total of 223 titles were identified, of which 129 were screened by full-text review. Eleven RCTs encompassing 5,847 patients met the inclusion criteria and were considered eligible for further analysis (−80 mm Hg: six studies; −125 mm Hg: five studies). A statistically significant improvement in the composite SSI (odds ratio [OR]: 0.69; 95% confidence interval [CI]: 0.54–0.89) and superficial SSI (OR: 0.66; 95% CI: 0.50–0.86) outcomes was observed with the −80 mm Hg device, compared with standard dressings. The same effect on SSI outcomes was not observed with the −125 mm Hg device (composite SSI—OR: 0.91; 95% CI: 0.64–1.28; superficial SSI—OR: 1.12; 95% CI: 0.70–1.78). There were no statistically significant differences in any of the other assessed outcomes. sNPWT devices may differ in their ability to reduce composite or superficial SSI after C-section.

Key Points

  • Negative pressure benefits obese patients undergoing C-section.

  • Negative pressure devices may differ in performance.

  • A head-to-head clinical trial is needed.


#

Deliveries by cesarean section (C-section) are one of the most common obstetric procedures worldwide, with rates increasing over the last 30 years.[1] [2] By 2030, it is projected that 28.5% of all births will be by C-section, with an even higher rate of 33.8% in the United States.[3] Brazil has one of the highest rates, globally, at 55.8%.[4]

The rising prevalence of C-section deliveries is driven by elements of perception (i.e., it is controllable and convenient[5]) and an increase in medical risk factors, especially prepregnancy obesity (body mass index [BMI] ≥ 30 kg/m2).[6] Prepregnancy obesity in the United States rose from 26.1% in 2016 to 29.0% in 2019.[7] Surgeon preference for incision types in obese patients is largely weighted towards using a Pfannenstiel incision,[8] although vertical incisions have been shown to have roughly similar postoperative wound morbidity in this population.[9]

The incidence of surgical site infection (SSI) after C-section is approximately 10% in the United States.[10] Prepregnancy obesity is an additional risk factor for SSI after C-section[11] [12] [13] [14] [15] and other surgical site complications (SSCs), including dehiscence, seroma, hematoma, and bleeding.[16] [17] [18]

Prophylactic negative pressure wound therapy (NPWT) has emerged as a successful intervention in patients undergoing C-section at high risk for SSI[19] and other SSCs. Single-use NPWT (sNPWT) devices are especially advantageous, as they are small, light, and highly portable.[20] The two most widely used sNPWT devices following C-section are PICO (Smith & Nephew Medical Ltd; Hull, United Kingdom;−80 mm Hg device) and PREVENA (previously KCI, an Acelity Company, now 3M; San Antonio, TX; −125 mm Hg device), which have been approved since 2010 by the U.S. Food and Drug Administration for prophylactic application after incisional closure at the time of surgery.[21]

Both consist of a single-use, battery-powered device and a foam-based or absorbent layer-based, peel-and-place dressing designed for closed surgical incisions. There have been multiple studies comparing each device against standard dressings, but none offering head-to-head comparisons of outcomes with these two devices in this indication.[22]

With the −80 mm Hg device, exudate is managed predominantly by evaporation through the dressing, therefore negating the need for a canister.[23] It is a multilayered adhesive dressing including an AIRLOCK (Smith & Nephew Medical Ltd; Hull, United Kingdom) Technology layer that delivers consistent negative pressure across the whole dressing to ensure treatment is delivered to a wider zone beyond the wound itself.[24] The device delivers continuous negative pressure at −80 mm Hg for up to 7 days.[23]

The −125 mm Hg device collects exudate in a replaceable 45-mL canister[25] and uses a reticulated polyurethane foam dressing covered with a thin film. The device delivers continuous negative pressure at −125 mm Hg for up to 7 days.[26]

Both devices are currently indicated in the United States to aid in the reduction of the incidence of postoperative seroma and, in patients at high risk for postoperative infection, superficial SSI in Class I and II wounds. The −80 mm Hg device is also additionally indicated in the United States to aid in the reduction of the incidence of dehiscence and deep SSI in Class I and II wounds.

Previous studies have shown inconsistencies in the clinical outcomes obtained with sNPWT devices, which may be in part due to differences in the sNPWT interventions used between studies. Several recent systematic literature reviews and meta-analyses[27] [28] [29] have demonstrated reductions in the incidence of SSIs by using sNPWT for the management of closed surgical incisions in patients undergoing C-section. However, most of these studies have only reported on the pooled data of the different sNPWT devices versus standard of care. Guo et al investigated different sNPWT devices as part of a secondary subgroup analysis but their study was not specifically designed to answer the question of whether differences in performance exist between sNPWT devices and confined their analysis to a single outcome (composite SSI).[28]

Therefore, the research question this study aims to answer is whether there are any differences in relation to standard care in the clinical performances of two sNPWT devices across a number of different wound complication outcomes. By answering the question, this analysis will expand on previous studies in this area by providing additional information on variables likely to influence successful wound healing in this patient population and guide future research.

Sources

This review was written in accordance with Preferred Reporting Items for Systematic Reviews and Meta-analyses (PRISMA) guidelines.[30] This review was not preregistered.

A search was undertaken using Embase, PubMed, Cochrane Library, and ClinicalTrials.gov to identify studies reporting on the use of either the −80 or the −125mm Hg devices. The search string/keywords can be found in the supplementary material ([Supplementary Table S1], available in online version). To increase the scope of the search, the search terms were left intentionally broad (e.g., no outcomes were specified).


#

Study Selection

English language studies published from 2011 to September 2021 were included. Two experienced data reviewers screened for relevant studies independently by examining titles and abstracts. To be included, the study must have been a randomized controlled trial (RCT) with ≥10 patients in each treatment arm and report on the use of sNPWT (−80 or −125 mm Hg device) for the management of surgical site incisions in patients of any age undergoing C-section. The study must also have reported on obese patients (BMI ≥ 30 kg/m2). RCTs deemed potentially relevant progressed to full-text screening. In instances of disagreement, a third reviewer made the final decision for inclusion after reading the full-text paper or conference abstract. Included RCTs compared outcomes following the use of either −80 or −125 mm Hg device versus standard care for closed surgical incisions in obstetrics (i.e., the use of standard non-NPWT dressings).

Data were extracted from included RCTs using a predefined and standardized data extraction form and checked for accuracy by a second reviewer. No automated extraction tools were used. Extracted data included descriptions of location of the RCT, the number of patients in each treatment arm, type of dressing used, treatment duration, follow-up period, and outcomes data. Outcome data were extracted based on the values reported at final assessment or follow-up. Where possible, conference abstracts were extracted in the same manner as full manuscripts. Missing or additional information from the abstract was obtained from the study's ClinicalTrials.gov registration.

The primary outcome of interest was the number of patients who had an SSI (classified as composite [overall], superficial, or deep) with either −80 or −125 mm Hg devices, compared with the standard of care. Secondary outcomes of interest were the number of patients who developed composite wound complications, dehiscence, seroma, hematoma, bleeding, or blistering. Readmission and reoperation rates were also collected.

The reviewers assessed the risk of bias for each RCT, considering the challenges of blinding clinicians and participants to sNPWT devices. Bias assessments were performed for individual RCTs using quality criteria taken from the Centre for Reviews and Dissemination guidelines[31] for the assessment of risk of bias in RCTs. Each criterion was rated for low, unclear/medium, and high risk of bias. To assess publication bias, an Egger's test was performed using the dmetar package in R 3.6.1 (R Core Team, 2019).

Overall effect estimates were calculated using the meta package in R. Formal meta-analyses were conducted when there were at least two RCTs reporting on the same outcome. The Mantel–Haenszel method was used for binary outcomes, with either a fixed effect model or a random effects model. Individual odds ratio (OR) estimates and summary estimates (including 95% confidence interval [CI]) were displayed graphically in forest plots. Heterogeneity was quantified with the I 2 statistic. The fixed effect model was used when heterogeneity was low (I 2 < 50%) and random effects model when heterogeneity was high (I 2 ≥ 50%). Sensitivity analyses were performed by the exclusion or inclusion of terminated RCTs and using fixed effect or random effects models. p < 0.05 denoted statistical significance.


#

Results

The PRISMA flow diagram outlining the study selection at various phases of the systematic literature review is shown in [Fig. 1].[32] In total, 223 studies were identified from the review, of which 129 were screened by full text. There was one study that appeared to meet the inclusion criteria but was excluded (i.e., “near-miss”), as it did not report the type of negative pressure device used.[33] No other “near-misses” were identified. Eleven RCTs met the inclusion criteria and were considered eligible for further analysis.

Zoom Image
Fig. 1 PRISMA flow diagram of included studies. PRISMA, Preferred Reporting Items for Systematic Reviews and Meta-Analyses.

Key study characteristics are outlined in [Table 1]. Six RCTs evaluated the −80 mm Hg device,[34] [35] [36] [37] [38] [39] of which two RCTs[36] [39] were available as conference abstracts only. Five RCTs evaluated the −125 mm Hg device.[25] [26] [40] [41] [42] Two RCTs, one reporting on the −80 mm Hg device and the other on the −125 mm Hg device, were identified as “terminated”.[38] [41] All studies reported on either composite, superficial, or deep SSI, except Gonzalez et al who only reported overall wound complications.[36] The majority of studies stated that all patients received antibiotic prophylaxis, except for some studies where their use was not mentioned.[34] [36] [37] The majority (>98%) of patients in the Gillespie et al's study received antibiotic prophylaxis.[35]

Table 1

Characteristics of included studies

Author, year, country, BMI inclusion criteria

Relevant outcome(s) assessed[b]

Antibiotic prophylaxis used? (Y/N/NR)

BMI (kg/m2)

Emergency C-section (frequency)

Final outcome assessment duration

Incisional dressings used

Subjects[c]

Treatment duration

SSI (overall)

SSI (superficial)

SSI (deep)

Wound complications

Dehiscence

Seroma

Hematoma

Bleeding

Blistering

Readmission

Reoperations

Hyldig et al 2018[37], Denmark,

BMI ≥ 30 kg/m2

x

x

x

x

NR

Median (IQR):

34.7 (31.5–38.2)

203 (47.1%)

30 days

−80 mm Hg device

432

5 days

Median (IQR):

34.2 (31.6–38.1)

209 (47.0%)

Standard postoperative dressing

444

The dressing was left in situ for at least 24 hours

Tuuli et al 2017[39], the United States,

BMI ≥ 30 kg/m2

x

x

x

x

x

Y

Mean (SD):

39.2 (9.2)

Included but not reported

30 days

−80 mm Hg device

60

Removed at discharge (usually on day 4)

Mean (SD):

40.7 (8.6)

Included but not reported

Standard wound dressing

60

The dressing was removed after 24 hours

Chaboyer et al 2014[34], Australia,

BMI ≥ 30 kg/m2

x

x

x

x

x

x

x

x

x

NR

Median (IQR): 35.7 (4.5)

Excluded

28 days

−80 mm Hg device

46

The dressing remained in place until day 4, unless it became soiled or dislodged, in that case a new dressing of the same type was applied

NR

Median (IQR): 36.8 (5.8)

Excluded

Standard dressing

46

The dressing remained in place until day 4, unless it became soiled or dislodged, in that case a new dressing of the same type was applied

Gillespie et al 2021[35], Australia,

BMI ≥ 30 kg/m2

x

x

x

x

x

x

x

x

x

x

x

Y[d]

BMI, n (%):

30–34.9: 488 (48)

35–39.9: 268 (26)

40–49.9: 218 (21)

≥50: 43 (4)

Excluded

30 days

−80 mm Hg device

1,017

The dressing was left intact for 5 to 7 days

BMI, n (%):

30 − 34.9: 524 (52)

35 − 39.9: 247 (24)

40 − 49.9: 211 (21)

≥50: 117 (12)

Excluded

Standard dressing

1018

The dressing was left intact for 5 to 7 days

Peterson et al 2021[a] [38], the United States,

BMI ≥ 40 kg/m2

x

x

x

x

x

x

Y

Mean (SD):

49.3 (6.6)

Unscheduled:

32 (58.2%)

6 weeks

−80 mm Hg device

55

The NPWT dressing was changed once by the discharging physician on postoperative day 3 or 4 to evaluate the wound before hospital discharge. If the indicator light on the negative-pressure dressing indicated that the dressing was no longer functioning, then the initial negative-pressure dressing was changed, and wound evaluation performed before postoperative day 3. The patient was given standardized verbal and written instructions on how to remove and discard the NPWT dressing on postoperative day 7

Mean (SD):

47.8 (6.9)

Unscheduled:

35 (63.6%)

Sterile nonadherent wound dressing, a sterile abdominal gauze pad, and a waterproof transparent adhesive dressing

55

The standard dressing was removed on postoperative day 1 and inspected daily during inpatient admission

Gonzalez et al 2020[36], the United States,

BMI ≥ 30 kg/m2

x

NR

Mean (SD):

39.9 (8.3)

NR

6 weeks

−80 mm Hg device

79

The NPWT dressing was removed on the day of hospital discharge and replaced with a second dressing. The patient removed the NPWT dressing on postoperative day 7

NR

Standard dressing

76

28 to 48 hours

Gunatilake et al 2017[26], the United States,

BMI ≥ 35 kg/m2

x

x

x

x

x

Y

Mean (SD):

46.3 (7.3)

Range:

35.7 − 60.8

NR

52 days

−125 mm Hg device

46

Duration of NPWT was ≥5 to ≤7 days, immediately following surgery

Mean (SD):

46.8 (5.6)

Range: 38.9 − 60.8

NR

Sterile adhesive strips, sterile gauze, and a waterproof transparent adhesive dressing

46

1 to 2 days

Hussamy et al 2019[25], United States,

BMI ≥ 40 kg/m2

x

x

x

x

x

x

x

Y

Mean (SD):

46.6 (6.0)

Urgent: 141 (64%)

Emergency: 9 (4%)

60 days

−125 mm Hg device

222

Until hospital discharge or unless premature removal was indicated or requested by the primary obstetric care team

Mean (SD):

45.8 (5.8)

Urgent: 138 (63%)

Emergency: 9 (4%)

Standard dressing

219

Per hospital routine, usually day 1 postoperatively

Ruhstaller et al 2017[40], the United States,

BMI ≥ 30 kg/m2

x

x

x

x

Y

Median (range):

36.1 (33.2 − 41.8)

Unscheduled:

61 (100%)

4 weeks

−125 mm Hg device

67

The dressing was removed at day 3 postoperatively

Median (range):

35.1 (32.6 − 42.1)

Unscheduled:

57 (98%)

Sterile nonadherent wound dressing overlaid with a 4 cm × 4 cm gauze pad and surgical tape

69

Dressing was removed 24 hours after surgery

Wihbey et al 2018[42], the United States,

BMI ≥ 35 kg/m2

x

x

x

x

x

x

x

x

x

Y

Mean (SD):

44.9 (8.0)

Scheduled:

9 (12%)

30 days

−125 mm Hg device

80

Dressing was removed between postoperative days 5 and 7

Mean (SD):

43.4 (7.0)

Scheduled: 7 (8%)

Sterile nonadherent wound dressing, a sterile gauze, and a waterproof transparent adhesive dressing

86

The dressing was removed on postoperative day 2

Tuuli et al 2020[a] [41], the United States,

BMI ≥ 30 kg/m2

x

x

x

x

x

x

x

x

x

x

Y

Mean (SD):

39.6 (7.7)

Unscheduled:

253 (31.4%)

Urgent:

90 (11.2%)

Emergency: 46 (5.7%)

30 days

−125 mm Hg device

816

Removed on discharge at postoperative day 4 or 7 for patients who remained hospitalized

Mean (SD):

39.5 (8.1)

Unscheduled:

256 (31.9%)

Urgent:

96 (12.0%)

Emergency: 36 (4.5%)

Standard dressing

808

Dressing was removed at 24 hours

Abbreviations: BMI, body mass index; C-section, cesarean section; IQR, interquartile range; NPWT, negative pressure wound therapy; NR, not reported; RCT, randomized controlled trial; SD, standard deviation; SSI, surgical site infection.


a Terminated RCT.


b Studies reported upon additional outcomes not relevant to this analysis.


c At randomization.


d The authors report that the majority of patients received antibiotic prophylaxis although not all patients received prophylaxis.


In total, 5,847 patients were investigated from the collective RCTs (numbers at randomization). For the −80 mm Hg device RCTs, 1,689 patients received the intervention and 1,699 patients received standard dressings. For the −125 mm Hg device RCTs, 1,231 patients received the intervention and 1,228 patients received standard dressings.

The results of the bias assessment for the included studies can be seen in [Fig. 2]. The overall quality of the studies was deemed acceptable, with the studies at low-to-moderate risk of bias. This was consistent between both −80 and −125 mm Hg device studies. The main factor contributing to increased risk of bias among most studies was an unclear, or unavailable, intention-to-treat analysis. No studies were completely blinded due to the nature of the interventions being administered.

Zoom Image
Fig. 2 Bias assessment of included studies. ITT, intention to treat; RCT, randomized controlled trial.

An overview of the meta-analysis results is shown in [Table 2]. A statistically significant improvement in the composite SSI (OR: 0.69; 95% CI: 0.54–0.89) and superficial SSI (OR: 0.66; 95% CI: 0.50–0.86) outcomes was observed with the −80 mm Hg device ([Figs. 3] and [4]), compared with standard of care. The same effect on SSI outcomes was not observed with the −125 mm Hg device (composite SSI—OR: 0.91; 95% CI: 0.64–1.28; superficial SSI—OR: 1.12; 95% CI: 0.70–1.78; [Figs. 5] and [6]). There were no statistically significant improvements in deep SSI outcomes for either the −80 mm Hg device (OR: 0.84; 95% CI: 0.43–1.66) or the −125 mm Hg device (OR: 0.99; 95% CI: 0.43–2.31).

Zoom Image
Fig. 3 Forest plot for the −80 mm Hg device; composite surgical site infection outcome. CI, confidence interval.
Zoom Image
Fig. 4 Forest plot for the −80 mm Hg device; superficial surgical site infection outcome.
Zoom Image
Fig. 5 Forest plot for the −125 mm Hg device; composite surgical site infection outcome.
Zoom Image
Fig. 6 Forest plot for the −125 mm Hg device; superficial surgical site infection outcome.
Table 2

Results of meta-analyses performed for all surgical site complication outcomes

Outcome

Device

Number of studies

Number of participants

Statistical method

I 2 statistic (%)

Effect estimate

p-Value

SSI (composite)

−80 mm Hg device

5

3,228

OR (M–H, fixed effect, 95% CI)

0

0.69 (0.54–0.89)

0.0043[a]

−125 mm Hg device

5

2,411

0

0.91 (0.64–1.28)

0.5867

Superficial SSI

−80 mm Hg device

4

3,108

OR (M–H, fixed effect, 95% CI)

22

0.66 (0.50–0.86)

0.0025[a]

−125 mm Hg device

3

1,848

0

1.12 (0.70–1.78)

0.6430

Deep SSI

−80 mm Hg device

3

2,998

OR (M–H, fixed effect, 95% CI)

0

0.84 (0.43–1.66)

0.6243

−125 mm Hg device

3

1,848

NA

0.99 (0.43–2.31)

0.9906

Wound complications (composite)[b]

−80 mm Hg device

4

472

OR (M–H, fixed effect, 95% CI)

0

0.85 (0.53–1.35)

0.4834

−125 mm Hg device

5

2,411

0

0.89 (0.68–1.17)

0.4017

Seroma

−80 mm Hg device

4

2,352

OR (M–H, fixed effect, 95% CI)

0

1.04 (0.62–1.75)

0.8915

−125 mm Hg device

3

1,851

0

0.82 (0.39–1.74)

0.6072

Dehiscence

−80 mm Hg device

4

3,069

OR (M–H, fixed effect, 95% CI)

0

1.01 (0.80–1.26)

0.9444

−125 mm Hg device

4

1,930

26.9

1.11 (0.65–1.90)

0.7074

Hematoma

−80 mm Hg device

3

2,265

OR (M–H, fixed effect, 95% CI)

0

1.94 (0.75–5.01)

0.1736

−125 mm Hg device

2

1,769

0

0.49 (0.18–1.33)

0.1623

Bleeding

−80 mm Hg device

2

2,122

OR (M–H, fixed effect, 95% CI)

0

0.88 (0.44–1.77)

0.7207

−125 mm Hg device

NA

NA

NA

NA

NA

NA

Reoperation

−80 mm Hg device

NA

NA

NA

NA

N/A

NA

−125 mm Hg device

4

803

OR (M–H, fixed effect, 95% CI)

14

1.00 (0.52–1.93)

0.9914

Readmission

−80 mm Hg device

3

2,232

OR (M–H, fixed effect, 95% CI)

0

1.68 (0.88–3.21)

0.1147

−125 mm Hg device

3

2,210

0

1.21 (0.59–2.46)

0.6008

Blistering

−80 mm Hg device

3

2,176

OR (M–H, random effect, 95% CI)

54

1.42 (0.40–5.04)

0.5922

−125 mm Hg device

2

1,727

OR (M–H, random effect, 95% CI)

70

12.71 (0.69–235.29)

0.0878

Abbreviations: CI, confidence interval; M–H, Mantel–Haenszel; NA, not available; OR, odds ratio; SSI, surgical site infection.


a Denotes statistical significance.


b The definition of wound complications differed between included studies but generally included seroma, dehiscence, hematoma, bleeding, and blistering.


There was no statistically significant difference in the composite wound complication outcome for either −80 mm Hg (OR: 0.85; 95% CI: 0.53–1.35) or −125 mm Hg (OR: 0.89; 95% CI: 0.68–1.17) interventions, compared to standard of care. Meta-analysis of each component of the composite wound complications outcome also demonstrated no statistically significant reduction in dehiscence (−80 mm Hg device—OR: 1.01; 95% CI: 0.80–1.26; −125 mm Hg device—OR: 1.11; 95% CI: 0.65–1.90) and seroma (−80 mm Hg device—OR: 1.04; 95% CI: 0.62–1.75; −125 mm Hg device—OR: 0.82; 95% CI: 0.39–1.74). All other SSCs analyzed (hematoma, bleeding) were also not statistically significant ([Table 2]; [Supplementary Figs. S1]–[S16]; available in the online version). No statistical difference was observed with the readmission (−80 mm Hg device—OR: 1.68; 95% CI: 0.88–3.21; −125 mm Hg device—OR: 1.21; 95% CI: 0.59–2.46) outcome. A meta-analysis for the reoperation outcome was not possible for the −80 mm Hg device due to insufficient data. No statistical difference was observed with the −125 mm Hg device compared to standard of care for this outcome (OR: 1.00; 95% CI: 0.52–1.93).

For the blistering outcome, no statistically significant increase was identified with the use of either the −80 mm Hg device (three RCTs; 1,095 patients in the intervention group and 1,081 patients in the control group; OR: 1.42; 95% CI: 0.40–5.04) or the −125 mm Hg device (two RCTs; 867 patients in the intervention group and 860 patients in the control group; OR: 12.71; 95% CI: 0.69–235.29) compared to standard of care. One study with the −125 mm Hg device was halted due to increased blistering.[41] An additional study reported a number of blistering cases in the −125 mm Hg group compared to zero in the standard group, although the authors pooled these events with other skin reactions making it unsuitable for inclusion in this subanalysis.[25]

Statistical heterogeneity denoted by the I 2 statistic was not significant (<50%) for all outcomes, except for the −80 and −125 mm Hg devices blistering outcome subanalyses (I 2 = 54% and I 2 = 70%, respectively). Sensitivity analyses were performed for the composite SSI and superficial SSI outcomes. Applying a random effects model, it was still possible to demonstrate a statistically significant reduction in composite SSI (OR: 0.70; 95% CI: 0.54–0.90) and superficial SSI (OR: 0.64; 95% CI: 0.44–0.93) for the −80 mm Hg device, but no statistically significant difference for the −125 mm Hg device when compared to standard of care. Further sensitivity analyses were performed by eliminating the terminated studies from each subanalysis to determine their impact on the results. Both the composite (OR: 0.68; 95% CI: 0.52–0.88) and superficial SSI (OR: 0.64; 95% CI: 0.48–0.85) outcomes remained statistically significant for the −80 mm Hg device, with no statistically significant difference detected with either outcome for the −125 mm Hg device.

The Egger's publication bias test did not indicate the presence of funnel plot asymmetry for the −80 mm Hg device composite SSI (intercept: 0.491; 95% CI: −1.08 to 2.06; t = 0.613; p-value = 0.5833) and superficial SSI outcomes (intercept: −0.387; 95% CI: −3.01 to 2.24; t = − 0.289; p-value = 0.7997). For details on the funnel plots, refer to the supplementary material ([Supplementary Figs. S17]–[S36], available in the online version).


#

Discussion

This systematic literature review identified a balanced number of RCTs investigating either −80 or −125 mm Hg devices for the management of surgical site incisions in patients undergoing C-section. It demonstrates a significant benefit of using sNPWT devices to reduce SSIs after C-section over standard dressings, which agrees with recent similar systematic literature reviews[27] [28] [29] as well as previous studies.[21] [43] Importantly, the present study goes further to demonstrate that this significant benefit in the reduction of SSI may be device specific.

These findings agree with a previous systematic literature review and meta-analysis conducted by Guo et al in 2021, who identified a difference between the two sNPWT devices in a subgroup analysis for overall SSI.[28] The primary research question of the present study was to compare the two negative pressure devices in detail for the most clinically relevant wound complications such as SSI, seroma, hematoma, and dehiscence. This differs from the study by Guo et al who only investigated device differences as part of a secondary analysis for a single outcome and were, therefore not capable of answering this question comprehensively. Specifically, Guo et al did not evaluate SSI subtypes (superficial, deep) or any other key wound complication outcome in relation to a device subgroup analysis. The present study, therefore, provides a detailed comparison of the clinical performances of the two devices, which allows for greater understanding of what variables contribute to successful clinical outcomes in obese patients undergoing C-section. In addition, the present updated analysis provides additional relevant data not included in previous reviews, thereby providing the most comprehensive picture to date in this research area.

The main finding of this analysis, therefore, expands on prior reviews in demonstrating that using the −80 mm Hg device resulted in a significantly lower number of superficial SSIs (as well as overall SSIs) compared with standard dressings. This significant difference was not demonstrated with the −125 mm Hg device (when compared to standard of care). Various sensitivity analyses performed on these outcomes demonstrate the robustness of these findings. Although the majority of the treatment effect on SSI reduction is being driven by a reduction in superficial SSIs, this is to be expected given the low relative incidence of deep SSIs, as most studies will not be sufficiently powered to detect a statistical difference.[44] A statistically significant reduction in superficial SSI is, therefore, clinically relevant as this subtype represents the majority of SSIs.

These differences could be explained with the differing mode of action of the devices. The −80 mm Hg device has a primary mode of action of evaporation through four layers whereas the −125 mm Hg device has a foam filler and a separate exudate canister. They also operate at different negative pressures; however, there have been no head-to-head studies to date that would have detected such a difference. Brownhill et al demonstrated that use of the −80 mm Hg device resulted in faster stimulation of reepithelialization and promotion of granulation tissue than using traditional NPWT devices, which generate negative pressure between −50 and −175 mm Hg.[45] Wilkinson et al reported that a heightened damage response in the epidermis was shown with a traditional NPWT device generating a negative pressure of −100 mm Hg compared to one of the devices in this systematic literature review which generates a pressure of −80 mm Hg.[46] This was in addition to the studies that have shown that retention of wound filler can also increase inflammatory cell influx and promote foreign body reactions.[47] [48]

The reduction of SSIs is an outcome of interest to patients, clinicians, and health care systems. SSI following C-section causes pain and anxiety to new parents, and incurs a financial burden to the health care system in both community and hospital health care settings.[49] An SSI can impact the quality of life of the gestational carrier and risk to bonding with the newborn.[50] The postoperative length of stay in a hospital setting is often increased,[51] along with higher rates of readmission and in some cases postpartum mortality.[52] SSIs in C-section have been estimated to cost approximately £1,866 per infection in the United Kingdom[49] and $3,500 per infection in the United States.[53] These complications are increasingly problematic in light of the rising global prevalence of obesity,[54] which is a known contributor to postoperative SSIs; therefore, considering therapies that may reduce this and other wound complications is important.[55]

There is a substantial price difference in the cost of each sNPWT device kit, with the −80 mm Hg device being the cheaper of the two devices. Using prices from the United Kingdom as an example, the −80 mm Hg device costs between $159 and $182,[56] while the −125 mm Hg device costs between $373 and $438.[57] From a U.S. payer perspective, the −80 mm Hg device has been reported to save $637 per patient when compared to standard care, with greater savings achievable in higher risk patients such as those with a BMI ≥ 30 kg/m2.[58] These findings are based on composite SSI incidence rates and so account for the disproportionate health care costs and differences in device efficacy associated with the various SSI subtypes. Clinicians should, therefore, be aware of the differences in clinical and economic outcomes obtained with various sNPWT devices and factor this into any decisions made regarding their use.

The studies included in this systematic review all had patients with a BMI ≥ 30 kg/m2. These patients are particularly at risk of SSC when undergoing C-section. Increasing BMI may be correlated with an increasing incidence of SSCs. For instance, those with even higher BMIs, surpassing 40 kg/m2, may be at increasingly higher risk compared to patients with a BMI of 30–35 kg/m2.[16] [59]

Neither the −80 nor the −125 mm Hg device demonstrated a benefit in any of the other assessed outcomes (e.g., wound complications, seroma, dehiscence), which is consistent with the Gillespie et al 2022, Guo et al, and Angarita et al reviews,[27] [28] [29] who found no difference in the pooled sNPWT data when comparing to standard care. This apparent lack of treatment effect could be because many studies are not powered to detect these complications in sufficient numbers as they occur at a relatively lower frequency than compared to SSI. Furthermore, these outcomes were not as frequently reported between studies, meaningless data were available for assessment. Further studies may or may not establish whether a statistically significant treatment effect can be obtained with these devices in other outcomes. Despite no currently identified difference in the other assessed outcomes, the use of sNPWT is justified on the basis of a reduction in SSIs alone due to the associated morbidity, mortality, and subsequent health care costs caused by this complication.

Previous studies[27] [28] [29] have noted that the incidence of blistering is higher with sNPWT dressings compared with standard dressings. In this systematic literature review, no statistically significant increase was shown in blistering with the −80 or the −125 mm Hg device. A trend toward increased incidence of blistering with both devices was identifiable, however. Stratifying the blistering outcome by sNPWT device results in insufficient data to detect a statistically significant difference and may be reflected in the moderate-to-high statistical heterogeneity observed for these subanalyses. The results of this study show that blistering is not a unique phenomenon associated with a particular type of sNPWT device but is a common issue when using this intervention more generally. The incidence of blistering should be considered in context of the clinical benefit obtained with using sNPWT, as a statistically significant reduction in the odds of developing SSIs is likely more desirable to clinicians than the occurrence of a minor, self-limiting adverse event. In terms of clinical practice, further product training for health care professionals may be required on application technique, which may help to reduce this blistering effect.[56]

Most studies were at risk of bias because of a lack of blinding; however, blinding is acknowledged to be very difficult due to the nature of these interventions. There could be variations in outcomes based on surgical approach taken (i.e., whether a Pfannenstiel or a vertical incision was made) but this was not apparent from the studies included in this systematic literature review. Additional variables between studies that may have influenced the results include the use of antibiotic prophylaxis and the proportion of pregnant patients undergoing emergency C-sections. Most studies reported the use of antibiotic prophylaxis, although some studies omitted this detail. A higher frequency of emergency C-sections would indicate a higher risk population and therefore higher baseline complication rate, making it easier to detect a statistical difference in the outcome of interest. Other notable differences between studies were the variation in follow-up times and treatment durations between each study, which will increase the degree of heterogeneity and result in wider CIs for each outcome. The length of treatment (sNPWT) duration is based on clinician's judgement and influenced by the patient's response to therapy. The sNPWT devices included in this analysis can be used for up to 7 days continuously. The employment of the more conservative random effects model was used in instances where heterogeneity was high and may account for some of this type of methodological heterogeneity. Future studies should aim to investigate the impact of these factors on the treatment effect observed with sNPWT devices.

Strict inclusion criteria focused on RCT-only data. This approach decreases heterogeneity and allows for narrower CIs in determining benefit in this patient population. However, it reduces external validity due to exclusion of observational studies, which are subject to higher degrees of bias but can account for a wider breadth of patient populations/risk factors. Inclusion of terminated RCTs can limit the influence of publication bias but can introduce other problems, such as the over- or underestimation of treatment effects, depending on the reasons for study termination (e.g., apparent treatment benefit or lack thereof). Specifically, one −80 mm Hg device study[38] was terminated early due to slow enrollment, resulting in performance of an unplanned interim analysis that showed a lack of treatment effect. Similarly, Tuuli et al 2020 terminated their study also due to a lack of identified treatment effect and an increase in adverse event rates in the intervention arm.[41] These two studies may, therefore, have led to an underestimation of the treatment effects for both interventions, although sensitivity analyses performed to account for this by removing these terminated studies did not alter the main findings. Overall, further studies are required to fully elucidate the treatment effect across all the relevant SSC outcomes. Specifically, a head-to-head trial directly comparing −80 to −125 mm Hg devices would provide additional clarity as to whether one device should be selected over another in this patient population, particularly for SSCs other than SSI.


#

Conclusion

The −80 mm Hg sNPWT device significantly reduces overall and superficial SSIs after C-section in obese pregnant patients compared to standard of care, an effect not observed with the −125 mm Hg device. More studies and larger patient numbers are needed to confirm the extent of the device-specific effect and understand the benefits of sNPWT in outcomes other than SSI.


#
#

Conflict of Interest

T.G. has a consultancy agreement with Smith & Nephew for this work and B.C. is an employee of Smith & Nephew.

Acknowledgments

The authors would like to thank Amy Glasswell, Christopher Saunders, and Andrea Karduss (Smith & Nephew, Inc.) for providing study screening, writing support, and reviewing the manuscript.

Supplementary Material

  • References

  • 1 Betran AP, Torloni MR, Zhang JJ, Gülmezoglu AM. WHO Working Group on Caesarean Section. WHO Statement on Caesarean Section Rates. BJOG 2016; 123 (05) 667-670
  • 2 Boerma T, Ronsmans C, Melesse DY. et al. Global epidemiology of use of and disparities in caesarean sections. Lancet 2018; 392 (10155): 1341-1348
  • 3 Betran AP, Ye J, Moller AB, Souza JP, Zhang J. Trends and projections of caesarean section rates: global and regional estimates. BMJ Glob Health 2021; 6 (06) e005671
  • 4 Rudey EL, Leal MDC, Rego G. Cesarean section rates in Brazil: trend analysis using the Robson classification system. Medicine (Baltimore) 2020; 99 (17) e19880
  • 5 Betrán AP, Temmerman M, Kingdon C. et al. Interventions to reduce unnecessary caesarean sections in healthy women and babies. Lancet 2018; 392 (10155): 1358-1368
  • 6 Kawakita T, Reddy UM, Landy HJ, Iqbal SN, Huang CC, Grantz KL. Indications for primary cesarean delivery relative to body mass index. Am J Obstet Gynecol 2016; 215 (04) 515.e1-515.e9
  • 7 Driscoll AK, Gregory ECW. Increases in prepregnancy obesity: United States, 2016–2019. NCHS Data Brief 2020; (392) 1-8
  • 8 Smid MC, Smiley SG, Schulkin J, Stamilio DM, Edwards RK, Stuebe AM. The Problem of the Pannus: Physician Preference Survey and a Review of the Literature on Cesarean Skin Incision in Morbidly Obese Women. Am J Perinatol 2016; 33 (05) 463-472
  • 9 Martin J, Gastanaduy M, Nakahara A. et al. Association of skin incision type with postoperative cesarean delivery complications in morbidly obese patients. Obstet Gynecol 2022; 139 (01) 14-20
  • 10 Shea SK, Soper DE. Prevention of cesarean delivery surgical site infections. Obstet Gynecol Surv 2019; 74 (02) 99-110
  • 11 Wloch C, Wilson J, Lamagni T, Harrington P, Charlett A, Sheridan E. Risk factors for surgical site infection following caesarean section in England: results from a multicentre cohort study. BJOG 2012; 119 (11) 1324-1333
  • 12 Anderson V, Chaboyer W, Gillespie B. The relationship between obesity and surgical site infections in women undergoing caesarean sections: an integrative review. Midwifery 2013; 29 (12) 1331-1338
  • 13 Marchi J, Berg M, Dencker A, Olander EK, Begley C. Risks associated with obesity in pregnancy, for the mother and baby: a systematic review of reviews. Obes Rev 2015; 16 (08) 621-638
  • 14 Moulton LJ, Eric Jelovsek J, Lachiewicz M, Chagin K, Goje O. A model to predict risk of postpartum infection after caesarean delivery. J Matern Fetal Neonatal Med 2018; 31 (18) 2409-2417
  • 15 Saadia Z. Association between maternal obesity and cesarean delivery complications. Cureus 2020; 12 (03) e7163
  • 16 Conner SN, Verticchio JC, Tuuli MG, Odibo AO, Macones GA, Cahill AG. Maternal obesity and risk of postcesarean wound complications. Am J Perinatol 2014; 31 (04) 299-304
  • 17 Ayres-de-Campos D. Obesity and the challenges of caesarean delivery: prevention and management of wound complications. Best Pract Res Clin Obstet Gynaecol 2015; 29 (03) 406-414
  • 18 Zaki MN, Truong M, Pyra M, Kominiarek MA, Irwin T. Wound complications in obese women after cesarean: a comparison of staples versus subcuticular suture. J Perinatol 2016; 36 (10) 819-822
  • 19 Searle RJ, Myers D. A survey of caesarean section surgical site infections with PICO™ Single Use Negative Pressure Wound Therapy System in high-risk patients in England and Ireland. J Hosp Infect 2017; 97 (02) 122-124
  • 20 Malmsjö M, Huddleston E, Martin R. Biological effects of a disposable, canisterless negative pressure wound therapy system. Eplasty 2014; 14: e15
  • 21 Yu L, Kronen RJ, Simon LE, Stoll CRT, Colditz GA, Tuuli MG. Prophylactic negative-pressure wound therapy after cesarean is associated with reduced risk of surgical site infection: a systematic review and meta-analysis. Am J Obstet Gynecol 2018; 218 (02) 200-210.e1
  • 22 Singh DP, Gabriel A, Silverman RP, Griffin LP, McGowan LD, D'Agostino Jr RB. Meta-analysis comparing outcomes of two different negative pressure therapy systems in closed incision management. Plast Reconstr Surg Glob Open 2019; 7 (06) e2259
  • 23 Hudson DA, Adams KG, Van Huyssteen A, Martin R, Huddleston EM. Simplified negative pressure wound therapy: clinical evaluation of an ultraportable, no-canister system. Int Wound J 2015; 12 (02) 195-201
  • 24 Saunders C, Nherera LM, Horner A, Trueman P. Single-use negative-pressure wound therapy versus conventional dressings for closed surgical incisions: systematic literature review and meta-analysis. BJS Open 2021; 5 (01) zraa003
  • 25 Hussamy DJ, Wortman AC, McIntire DD, Leveno KJ, Casey BM, Roberts SW. Closed incision negative pressure therapy in morbidly obese women undergoing cesarean delivery: a randomized controlled trial. Obstet Gynecol 2019; 134 (04) 781-789
  • 26 Gunatilake RP, Swamy GK, Brancazio LR. et al. Closed-incision negative-pressure therapy in obese patients undergoing cesarean delivery: a randomized controlled trial. AJP Rep 2017; 7 (03) e151-e157
  • 27 Gillespie BM, Thalib L, Ellwood D. et al. Effect of negative-pressure wound therapy on wound complications in obese women after caesarean birth: a systematic review and meta-analysis. BJOG 2022; 129 (02) 196-207
  • 28 Guo C, Cheng T, Li J. Prophylactic negative pressure wound therapy on surgical site infection in obese women after cesarean section: a systematic review and meta-analysis. Int J Gynaecol Obstet 2021; 158 (03) 502-511
  • 29 Angarita AM, Jayakumaran J, Di Mascio D, Berghella V. Prophylactic negative pressure wound therapy on wound complications after cesarean delivery in women with obesity: a meta-analysis of randomized controlled trials. Am J Obstet Gynecol MFM 2022; 4 (03) 100617
  • 30 Page MJ, McKenzie JE, Bossuyt PM. et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ 2021; 372 (71) n71
  • 31 Akers J. Systematic reviews: CRD's guidance for undertaking reviews in health care. York, United Kingdom: Centre for Reviews and Dissemination, University of York; 2009
  • 32 Haddaway NR, Page MJ, Pritchard CC, McGuinness LA. PRISMA2020: an R package and Shiny app for producing PRISMA 2020-compliant flow diagrams, with interactivity for optimised digital transparency and Open Synthesis. Campbell Syst Rev 2022; 18 (02) e1230
  • 33 Stitely M. Prevention of wound complications after cesarean delivery in obese women utilizing negative pressure wound therapy. Accessed June 9, 2023 at: https://clinicaltrials.gov/ct2/show/results/NCT00654641
  • 34 Chaboyer W, Anderson V, Webster J, Sneddon A, Thalib L, Gillespie BM. Negative pressure wound therapy on surgical site infections in women undergoing elective caesarean sections: a pilot RCT. Healthcare (Basel) 2014; 2 (04) 417-428
  • 35 Gillespie BM, Webster J, Ellwood D. et al. Closed incision negative pressure wound therapy versus standard dressings in obese women undergoing caesarean section: multicentre parallel group randomised controlled trial. BMJ 2021; 373 (893) n893
  • 36 Gonzalez MG, Elisa Barske M, Kjellsson KB, Saboda K, Hill MG. Topical negative pressure wound therapy to prevent wound complications following cesarean delivery in high risk obstetric patients. Reprod Sci 2020; 27 (01) 136A-137A
  • 37 Hyldig N, Vinter CA, Kruse M. et al. Prophylactic incisional negative pressure wound therapy reduces the risk of surgical site infection after caesarean section in obese women: a pragmatic randomised clinical trial. BJOG 2019; 126 (05) 628-635
  • 38 Peterson AT, Bakaysa SL, Driscoll JM, Kalyanaraman R, House MD. Randomized controlled trial of single-use negative-pressure wound therapy dressings in morbidly obese patients undergoing cesarean delivery. Am J Obstet Gynecol MFM 2021; 3 (05) 100410
  • 39 Tuuli MG, Martin S, Stout MJ. et al. 412: Pilot randomized trial of prophylactic negative pressure wound therapy in obese women after cesarean delivery. Am J Obstet Gynecol 2017; 216 (1, Supplement): S245
  • 40 Ruhstaller K, Downes KL, Chandrasekaran S, Srinivas S, Durnwald C. Prophylactic wound vacuum therapy after cesarean section to prevent wound complications in the obese population: a randomized controlled trial (the ProVac Study). Am J Perinatol 2017; 34 (11) 1125-1130
  • 41 Tuuli MG, Liu J, Tita ATN. et al. Effect of prophylactic negative pressure wound therapy vs standard wound dressing on surgical-site infection in obese women after cesarean delivery: a randomized clinical trial. JAMA 2020; 324 (12) 1180-1189 DOI: 10.1001/jama.2020.13361.
  • 42 Wihbey KA, Joyce EM, Spalding ZT. et al. Prophylactic negative pressure wound therapy and wound complication after cesarean delivery in women with Class II or III obesity: a randomized controlled trial. Obstet Gynecol 2018; 132 (02) 377-384 DOI: 10.1097/aog.0000000000002744.
  • 43 Huang HP, Zhao WJ, Pu J, He F. Prophylactic negative pressure wound therapy for surgical site infection in obese women undergoing cesarean section: an evidence synthesis with trial sequential analysis. J Matern Fetal Neonatal Med 2021; 34 (15) 2498-2505
  • 44 Ferraro F, Piselli P, Pittalis S. et al; Centro di Riferimento per le Infezioni associate alle Pratiche Assistenziali Latium Region Group. Surgical site infection after caesarean section: space for post-discharge surveillance improvements and reliable comparisons. New Microbiol 2016; 39 (02) 134-138
  • 45 Brownhill VR, Huddleston E, Bell A. et al. Pre-clinical assessment of single-use negative pressure wound therapy during in vivo porcine wound healing. Adv Wound Care (New Rochelle) 2021; 10 (07) 345-356
  • 46 Wilkinson HN, Longhorne FL, Roberts ER, Brownhill VR, Hardman MJ. Cellular benefits of single-use negative pressure wound therapy demonstrated in a novel ex vivo human skin wound model. Wound Repair Regen 2021; 29 (02) 298-305
  • 47 Dessy LA, Serratore F, Corrias F, Parisi P, Mazzocchi M, Carlesimo B. Retention of polyurethane foam fragments during VAC therapy: a complication to be considered. Int Wound J 2015; 12 (02) 132-136
  • 48 Mazoch M, Montgomery C. Retained wound vacuum foam in non-healing wounds: a real possibility. J Wound Care 2015; 24 (06) S18-S20
  • 49 Wloch C, Van Hoek AJ, Green N. et al. Cost-benefit analysis of surveillance for surgical site infection following caesarean section. BMJ Open 2020; 10 (07) e036919
  • 50 Bullough L, Burns S, Timmons J, Truman P, Megginson S. Reducing C-section wound complications. Clin Serv J 2015; ; (April): 2-6
  • 51 Blumenfeld YJ, El-Sayed YY, Lyell DJ, Nelson LM, Butwick AJ. Risk factors for prolonged postpartum length of stay following cesarean delivery. Am J Perinatol 2015; 32 (09) 825-832
  • 52 Zuarez-Easton S, Zafran N, Garmi G, Salim R. Postcesarean wound infection: prevalence, impact, prevention, and management challenges. Int J Womens Health 2017; 9: 81-88
  • 53 Olsen MA, Butler AM, Willers DM, Gross GA, Hamilton BH, Fraser VJ. Attributable costs of surgical site infection and endometritis after low transverse cesarean delivery. Infect Control Hosp Epidemiol 2010; 31 (03) 276-282
  • 54 Strauss A. Obesity in pregnant women: maternal, fetal, and transgenerational consequences. Eur J Clin Nutr 2021; 75 (12) 1681-1683
  • 55 Heslehurst N, Simpson H, Ells LJ. et al. The impact of maternal BMI status on pregnancy outcomes with immediate short-term obstetric resource implications: a meta-analysis. Obes Rev 2008; 9 (06) 635-683
  • 56 National Institute for Health and Care Excellence (NICE). PICO negative pressure wound dressings for closed surgical incisions: Medical technologies guidance [MTG43]. Accessed June 9, 2023 at: https://www.nice.org.uk/guidance/mtg43
  • 57 National Institute for Health and Care Excellence (NICE). Prevena incision management system for closed surgical incisions: Medtech innovation briefing [MIB173]. Accessed June 9, 2023 at: https://www.nice.org.uk/advice/mib173
  • 58 Nherera LM, Saunders C, Verma S, Trueman P, Fatoye F. Single-use negative pressure wound therapy reduces costs in closed surgical incisions: UK and US economic evaluation. J Wound Care 2021; 30 (Sup5): S23-S31
  • 59 Leth RA, Uldbjerg N, Nørgaard M, Møller JK, Thomsen RW. Obesity, diabetes, and the risk of infections diagnosed in hospital and post-discharge infections after cesarean section: a prospective cohort study. Acta Obstet Gynecol Scand 2011; 90 (05) 501-509

Address for correspondence

Ben Costa, MSc
Global Clinical and Medical Affairs, Smith + Nephew
101 Hessle Road, Hull HU3 2BN
United Kingdom   

Publication History

Received: 21 September 2022

Accepted: 17 August 2023

Article published online:
19 September 2023

© 2023. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA

  • References

  • 1 Betran AP, Torloni MR, Zhang JJ, Gülmezoglu AM. WHO Working Group on Caesarean Section. WHO Statement on Caesarean Section Rates. BJOG 2016; 123 (05) 667-670
  • 2 Boerma T, Ronsmans C, Melesse DY. et al. Global epidemiology of use of and disparities in caesarean sections. Lancet 2018; 392 (10155): 1341-1348
  • 3 Betran AP, Ye J, Moller AB, Souza JP, Zhang J. Trends and projections of caesarean section rates: global and regional estimates. BMJ Glob Health 2021; 6 (06) e005671
  • 4 Rudey EL, Leal MDC, Rego G. Cesarean section rates in Brazil: trend analysis using the Robson classification system. Medicine (Baltimore) 2020; 99 (17) e19880
  • 5 Betrán AP, Temmerman M, Kingdon C. et al. Interventions to reduce unnecessary caesarean sections in healthy women and babies. Lancet 2018; 392 (10155): 1358-1368
  • 6 Kawakita T, Reddy UM, Landy HJ, Iqbal SN, Huang CC, Grantz KL. Indications for primary cesarean delivery relative to body mass index. Am J Obstet Gynecol 2016; 215 (04) 515.e1-515.e9
  • 7 Driscoll AK, Gregory ECW. Increases in prepregnancy obesity: United States, 2016–2019. NCHS Data Brief 2020; (392) 1-8
  • 8 Smid MC, Smiley SG, Schulkin J, Stamilio DM, Edwards RK, Stuebe AM. The Problem of the Pannus: Physician Preference Survey and a Review of the Literature on Cesarean Skin Incision in Morbidly Obese Women. Am J Perinatol 2016; 33 (05) 463-472
  • 9 Martin J, Gastanaduy M, Nakahara A. et al. Association of skin incision type with postoperative cesarean delivery complications in morbidly obese patients. Obstet Gynecol 2022; 139 (01) 14-20
  • 10 Shea SK, Soper DE. Prevention of cesarean delivery surgical site infections. Obstet Gynecol Surv 2019; 74 (02) 99-110
  • 11 Wloch C, Wilson J, Lamagni T, Harrington P, Charlett A, Sheridan E. Risk factors for surgical site infection following caesarean section in England: results from a multicentre cohort study. BJOG 2012; 119 (11) 1324-1333
  • 12 Anderson V, Chaboyer W, Gillespie B. The relationship between obesity and surgical site infections in women undergoing caesarean sections: an integrative review. Midwifery 2013; 29 (12) 1331-1338
  • 13 Marchi J, Berg M, Dencker A, Olander EK, Begley C. Risks associated with obesity in pregnancy, for the mother and baby: a systematic review of reviews. Obes Rev 2015; 16 (08) 621-638
  • 14 Moulton LJ, Eric Jelovsek J, Lachiewicz M, Chagin K, Goje O. A model to predict risk of postpartum infection after caesarean delivery. J Matern Fetal Neonatal Med 2018; 31 (18) 2409-2417
  • 15 Saadia Z. Association between maternal obesity and cesarean delivery complications. Cureus 2020; 12 (03) e7163
  • 16 Conner SN, Verticchio JC, Tuuli MG, Odibo AO, Macones GA, Cahill AG. Maternal obesity and risk of postcesarean wound complications. Am J Perinatol 2014; 31 (04) 299-304
  • 17 Ayres-de-Campos D. Obesity and the challenges of caesarean delivery: prevention and management of wound complications. Best Pract Res Clin Obstet Gynaecol 2015; 29 (03) 406-414
  • 18 Zaki MN, Truong M, Pyra M, Kominiarek MA, Irwin T. Wound complications in obese women after cesarean: a comparison of staples versus subcuticular suture. J Perinatol 2016; 36 (10) 819-822
  • 19 Searle RJ, Myers D. A survey of caesarean section surgical site infections with PICO™ Single Use Negative Pressure Wound Therapy System in high-risk patients in England and Ireland. J Hosp Infect 2017; 97 (02) 122-124
  • 20 Malmsjö M, Huddleston E, Martin R. Biological effects of a disposable, canisterless negative pressure wound therapy system. Eplasty 2014; 14: e15
  • 21 Yu L, Kronen RJ, Simon LE, Stoll CRT, Colditz GA, Tuuli MG. Prophylactic negative-pressure wound therapy after cesarean is associated with reduced risk of surgical site infection: a systematic review and meta-analysis. Am J Obstet Gynecol 2018; 218 (02) 200-210.e1
  • 22 Singh DP, Gabriel A, Silverman RP, Griffin LP, McGowan LD, D'Agostino Jr RB. Meta-analysis comparing outcomes of two different negative pressure therapy systems in closed incision management. Plast Reconstr Surg Glob Open 2019; 7 (06) e2259
  • 23 Hudson DA, Adams KG, Van Huyssteen A, Martin R, Huddleston EM. Simplified negative pressure wound therapy: clinical evaluation of an ultraportable, no-canister system. Int Wound J 2015; 12 (02) 195-201
  • 24 Saunders C, Nherera LM, Horner A, Trueman P. Single-use negative-pressure wound therapy versus conventional dressings for closed surgical incisions: systematic literature review and meta-analysis. BJS Open 2021; 5 (01) zraa003
  • 25 Hussamy DJ, Wortman AC, McIntire DD, Leveno KJ, Casey BM, Roberts SW. Closed incision negative pressure therapy in morbidly obese women undergoing cesarean delivery: a randomized controlled trial. Obstet Gynecol 2019; 134 (04) 781-789
  • 26 Gunatilake RP, Swamy GK, Brancazio LR. et al. Closed-incision negative-pressure therapy in obese patients undergoing cesarean delivery: a randomized controlled trial. AJP Rep 2017; 7 (03) e151-e157
  • 27 Gillespie BM, Thalib L, Ellwood D. et al. Effect of negative-pressure wound therapy on wound complications in obese women after caesarean birth: a systematic review and meta-analysis. BJOG 2022; 129 (02) 196-207
  • 28 Guo C, Cheng T, Li J. Prophylactic negative pressure wound therapy on surgical site infection in obese women after cesarean section: a systematic review and meta-analysis. Int J Gynaecol Obstet 2021; 158 (03) 502-511
  • 29 Angarita AM, Jayakumaran J, Di Mascio D, Berghella V. Prophylactic negative pressure wound therapy on wound complications after cesarean delivery in women with obesity: a meta-analysis of randomized controlled trials. Am J Obstet Gynecol MFM 2022; 4 (03) 100617
  • 30 Page MJ, McKenzie JE, Bossuyt PM. et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ 2021; 372 (71) n71
  • 31 Akers J. Systematic reviews: CRD's guidance for undertaking reviews in health care. York, United Kingdom: Centre for Reviews and Dissemination, University of York; 2009
  • 32 Haddaway NR, Page MJ, Pritchard CC, McGuinness LA. PRISMA2020: an R package and Shiny app for producing PRISMA 2020-compliant flow diagrams, with interactivity for optimised digital transparency and Open Synthesis. Campbell Syst Rev 2022; 18 (02) e1230
  • 33 Stitely M. Prevention of wound complications after cesarean delivery in obese women utilizing negative pressure wound therapy. Accessed June 9, 2023 at: https://clinicaltrials.gov/ct2/show/results/NCT00654641
  • 34 Chaboyer W, Anderson V, Webster J, Sneddon A, Thalib L, Gillespie BM. Negative pressure wound therapy on surgical site infections in women undergoing elective caesarean sections: a pilot RCT. Healthcare (Basel) 2014; 2 (04) 417-428
  • 35 Gillespie BM, Webster J, Ellwood D. et al. Closed incision negative pressure wound therapy versus standard dressings in obese women undergoing caesarean section: multicentre parallel group randomised controlled trial. BMJ 2021; 373 (893) n893
  • 36 Gonzalez MG, Elisa Barske M, Kjellsson KB, Saboda K, Hill MG. Topical negative pressure wound therapy to prevent wound complications following cesarean delivery in high risk obstetric patients. Reprod Sci 2020; 27 (01) 136A-137A
  • 37 Hyldig N, Vinter CA, Kruse M. et al. Prophylactic incisional negative pressure wound therapy reduces the risk of surgical site infection after caesarean section in obese women: a pragmatic randomised clinical trial. BJOG 2019; 126 (05) 628-635
  • 38 Peterson AT, Bakaysa SL, Driscoll JM, Kalyanaraman R, House MD. Randomized controlled trial of single-use negative-pressure wound therapy dressings in morbidly obese patients undergoing cesarean delivery. Am J Obstet Gynecol MFM 2021; 3 (05) 100410
  • 39 Tuuli MG, Martin S, Stout MJ. et al. 412: Pilot randomized trial of prophylactic negative pressure wound therapy in obese women after cesarean delivery. Am J Obstet Gynecol 2017; 216 (1, Supplement): S245
  • 40 Ruhstaller K, Downes KL, Chandrasekaran S, Srinivas S, Durnwald C. Prophylactic wound vacuum therapy after cesarean section to prevent wound complications in the obese population: a randomized controlled trial (the ProVac Study). Am J Perinatol 2017; 34 (11) 1125-1130
  • 41 Tuuli MG, Liu J, Tita ATN. et al. Effect of prophylactic negative pressure wound therapy vs standard wound dressing on surgical-site infection in obese women after cesarean delivery: a randomized clinical trial. JAMA 2020; 324 (12) 1180-1189 DOI: 10.1001/jama.2020.13361.
  • 42 Wihbey KA, Joyce EM, Spalding ZT. et al. Prophylactic negative pressure wound therapy and wound complication after cesarean delivery in women with Class II or III obesity: a randomized controlled trial. Obstet Gynecol 2018; 132 (02) 377-384 DOI: 10.1097/aog.0000000000002744.
  • 43 Huang HP, Zhao WJ, Pu J, He F. Prophylactic negative pressure wound therapy for surgical site infection in obese women undergoing cesarean section: an evidence synthesis with trial sequential analysis. J Matern Fetal Neonatal Med 2021; 34 (15) 2498-2505
  • 44 Ferraro F, Piselli P, Pittalis S. et al; Centro di Riferimento per le Infezioni associate alle Pratiche Assistenziali Latium Region Group. Surgical site infection after caesarean section: space for post-discharge surveillance improvements and reliable comparisons. New Microbiol 2016; 39 (02) 134-138
  • 45 Brownhill VR, Huddleston E, Bell A. et al. Pre-clinical assessment of single-use negative pressure wound therapy during in vivo porcine wound healing. Adv Wound Care (New Rochelle) 2021; 10 (07) 345-356
  • 46 Wilkinson HN, Longhorne FL, Roberts ER, Brownhill VR, Hardman MJ. Cellular benefits of single-use negative pressure wound therapy demonstrated in a novel ex vivo human skin wound model. Wound Repair Regen 2021; 29 (02) 298-305
  • 47 Dessy LA, Serratore F, Corrias F, Parisi P, Mazzocchi M, Carlesimo B. Retention of polyurethane foam fragments during VAC therapy: a complication to be considered. Int Wound J 2015; 12 (02) 132-136
  • 48 Mazoch M, Montgomery C. Retained wound vacuum foam in non-healing wounds: a real possibility. J Wound Care 2015; 24 (06) S18-S20
  • 49 Wloch C, Van Hoek AJ, Green N. et al. Cost-benefit analysis of surveillance for surgical site infection following caesarean section. BMJ Open 2020; 10 (07) e036919
  • 50 Bullough L, Burns S, Timmons J, Truman P, Megginson S. Reducing C-section wound complications. Clin Serv J 2015; ; (April): 2-6
  • 51 Blumenfeld YJ, El-Sayed YY, Lyell DJ, Nelson LM, Butwick AJ. Risk factors for prolonged postpartum length of stay following cesarean delivery. Am J Perinatol 2015; 32 (09) 825-832
  • 52 Zuarez-Easton S, Zafran N, Garmi G, Salim R. Postcesarean wound infection: prevalence, impact, prevention, and management challenges. Int J Womens Health 2017; 9: 81-88
  • 53 Olsen MA, Butler AM, Willers DM, Gross GA, Hamilton BH, Fraser VJ. Attributable costs of surgical site infection and endometritis after low transverse cesarean delivery. Infect Control Hosp Epidemiol 2010; 31 (03) 276-282
  • 54 Strauss A. Obesity in pregnant women: maternal, fetal, and transgenerational consequences. Eur J Clin Nutr 2021; 75 (12) 1681-1683
  • 55 Heslehurst N, Simpson H, Ells LJ. et al. The impact of maternal BMI status on pregnancy outcomes with immediate short-term obstetric resource implications: a meta-analysis. Obes Rev 2008; 9 (06) 635-683
  • 56 National Institute for Health and Care Excellence (NICE). PICO negative pressure wound dressings for closed surgical incisions: Medical technologies guidance [MTG43]. Accessed June 9, 2023 at: https://www.nice.org.uk/guidance/mtg43
  • 57 National Institute for Health and Care Excellence (NICE). Prevena incision management system for closed surgical incisions: Medtech innovation briefing [MIB173]. Accessed June 9, 2023 at: https://www.nice.org.uk/advice/mib173
  • 58 Nherera LM, Saunders C, Verma S, Trueman P, Fatoye F. Single-use negative pressure wound therapy reduces costs in closed surgical incisions: UK and US economic evaluation. J Wound Care 2021; 30 (Sup5): S23-S31
  • 59 Leth RA, Uldbjerg N, Nørgaard M, Møller JK, Thomsen RW. Obesity, diabetes, and the risk of infections diagnosed in hospital and post-discharge infections after cesarean section: a prospective cohort study. Acta Obstet Gynecol Scand 2011; 90 (05) 501-509

Zoom Image
Fig. 1 PRISMA flow diagram of included studies. PRISMA, Preferred Reporting Items for Systematic Reviews and Meta-Analyses.
Zoom Image
Fig. 2 Bias assessment of included studies. ITT, intention to treat; RCT, randomized controlled trial.
Zoom Image
Fig. 3 Forest plot for the −80 mm Hg device; composite surgical site infection outcome. CI, confidence interval.
Zoom Image
Fig. 4 Forest plot for the −80 mm Hg device; superficial surgical site infection outcome.
Zoom Image
Fig. 5 Forest plot for the −125 mm Hg device; composite surgical site infection outcome.
Zoom Image
Fig. 6 Forest plot for the −125 mm Hg device; superficial surgical site infection outcome.