Endoscopy 2019; 51(05): 468-471
DOI: 10.1055/a-0656-5622
Innovations and brief communications
© Georg Thieme Verlag KG Stuttgart · New York

Clinical utility of novel ultrathin single-balloon enteroscopy: a feasibility study

Kaoru Takabayashi
1   Center for Diagnostic and Therapeutic Endoscopy, School of Medicine, Keio University, Tokyo, Japan
,
Naoki Hosoe
1   Center for Diagnostic and Therapeutic Endoscopy, School of Medicine, Keio University, Tokyo, Japan
,
Ryoichi Miyanaga
2   Division of Internal Medicine, Keio University School of Medicine, Tokyo, Japan
,
Seiichiro Fukuhara
1   Center for Diagnostic and Therapeutic Endoscopy, School of Medicine, Keio University, Tokyo, Japan
,
Kayoko Kimura
1   Center for Diagnostic and Therapeutic Endoscopy, School of Medicine, Keio University, Tokyo, Japan
,
Shinta Mizuno
2   Division of Internal Medicine, Keio University School of Medicine, Tokyo, Japan
,
Makoto Naganuma
2   Division of Internal Medicine, Keio University School of Medicine, Tokyo, Japan
,
Naohisa Yahagi
3   Cancer Center, School of Medicine, Keio University, Tokyo, Japan
,
Haruhiko Ogata
1   Center for Diagnostic and Therapeutic Endoscopy, School of Medicine, Keio University, Tokyo, Japan
,
Takanori Kanai
2   Division of Internal Medicine, Keio University School of Medicine, Tokyo, Japan
› Author Affiliations
Further Information

Publication History

submitted 09 April 2018

accepted after revision 21 June 2018

Publication Date:
01 August 2018 (online)

Abstract

Background The advent of balloon-assisted enteroscopy (BAE) has facilitated the examination of the entire digestive tract. However, using a rigid sliding tube during the procedure reduces patient acceptance. This study evaluated the clinical application of a newly developed ultrathin single-balloon enteroscope for BAE.

Methods 28 outpatients underwent enteroscopy with a novel ultrathin single-balloon enteroscope. None of the subjects required therapeutic procedures, such as balloon dilation or hemostasis. The insertability, efficacy, and safety of the ultrathin single-balloon endoscope were evaluated retrospectively.

Results 7 patients underwent transoral enteroscopy and 21 patients underwent transanal enteroscopy under conscious sedation. No adverse events related to the procedure were reported in any patients. Targeted observation and/or targeted biopsy were achieved in all procedures. All transoral procedures allowed evaluation of the jejunum beyond the ligament of Treitz. All transanal procedures allowed intubation of the terminal ileum, despite several patients having severe stenosis of the colon and ileum.

Conclusion A novel ultrathin single-balloon enteroscope showed adequate insertability and safety for outpatient surveillance enteroscopy under conscious sedation.

 
  • References

  • 1 Yamamoto H, Sekine Y, Sato Y. et al. Total enteroscopy with a nonsurgical steerable double-balloon method. Gastrointest Endosc 2001; 53: 216-220
  • 2 Tsujikawa T, Saitoh Y, Andoh A. et al. Novel single-balloon enteroscopy for diagnosis and treatment of the small intestine: preliminary experiences. Endoscopy 2008; 40: 11-15
  • 3 Rondonotti E, Spada C, Adler S. et al. Small-bowel capsule endoscopy and device-assisted enteroscopy for diagnosis and treatment of small-bowel disorders: European Society of Gastrointestinal Endoscopy (ESGE) Technical Review. Endoscopy 2018; 50: 423-446
  • 4 Sethi S, Thaker AM, Cohen J. et al. Monitored anesthesia care without endotracheal intubation is safe and efficacious for single-balloon enteroscopy. Dig Dis Sci 2014; 59: 2184-2190
  • 5 Atar M, Kadayifci A. Transnasal endoscopy: Technical considerations, advantages and limitations. World J Gastrointest Endosc 2014; 6: 41-48
  • 6 Sato K, Ito S, Shigiyama F. et al. A prospective randomized study on the benefits of a new small-caliber colonoscope. Endoscopy 2012; 44: 746-753
  • 7 Ohtsuka K, Kashida H, Kodama K. et al. Diagnosis and treatment of small bowel diseases with a newly developed single balloon endoscope. Dig Endosc 2008; 20: 134-137
  • 8 May A, Nachbar L, Schneider M. et al. Push-and-pull enteroscopy using the double-balloon technique: method of assessing depth of insertion and training of the enteroscopy technique using the Erlangen Endo-Trainer. Endoscopy 2005; 37: 66-70
  • 9 Upchurch BR, Sanaka MR, Lopez AR. et al. The clinical utility of single-balloon enteroscopy: a single-center experience of 172 procedures. Gastrointest Endosc 2010; 71: 1218-1223
  • 10 Horiuchi A, Nakayama Y. Unsedated ultrathin EGD by using a 5.2-mm-diameter videoscope: evaluation of acceptability and diagnostic accuracy. Gastrointest Endosc 2006; 64: 868-873
  • 11 Ai ZL, Lan CH, Fan LL. et al. Unsedated transnasal upper gastrointestinal endoscopy has favorable diagnostic effectiveness, cardiopulmonary safety, and patient satisfaction compared with conventional or sedated endoscopy. Surg Endosc 2012; 26: 3565-3572
  • 12 Tatsumi Y, Harada A, Matsumoto T. et al. Current status and evaluation of transnasal esophagogastroduodenoscopy. Dig Endosc 2009; 21: 141-146