Endoscopy 2002; 34(11): 905-908
DOI: 10.1055/s-2002-35302
Original Article
© Georg Thieme Verlag Stuttgart · New York

Evaluation of a New Three-Dimensional Magnetic Imaging System for Use During Colonoscopy

K.  Wehrmann1 , P.  Frühmorgen1
  • 1Dept. of Medicine, Ludwigsburg Hospital, Ludwigsburg, Germany
Further Information

Publication History

Submitted: 19 January 2002

Accepted after Revision: 30 July 2002

Publication Date:
13 November 2002 (online)

Background and Study Aims: A prototype magnetic imaging system (Scope Guide, Olympus Optical Co., Ltd.) provides a new facility for continuous viewing on a monitor of the position of the colonoscope during examination, without exposing patients or medical staff to radiation. The aim of this prospective study was to compare this magnetic imaging system with routine colonoscopy, including fluoroscopy. The study parameters were the detection of loops, the location of the endoscope tip at defined positions, the insertion time, and the premedication rate.
Materials and Methods: In the first part of the study, 133 consecutive patients were examined - 64 using an integrated three-dimensional colonoscope and 69 with the three-dimensional probe inserted into the biopsy channel of a routine video colonoscope. Fluoroscopy was used in all investigations for comparison at defined anatomical points and loops, and pathological findings and defined anatomic structures were documented using a laser printer both for three-dimensional colonoscopy and fluoroscopy. In the second part of the study, 25 further patients underwent colonoscopy with a modified prototype, now exclusively using the integrated three-dimensional colonoscope.
Results: The total time for insertion and the premedication rate did not differ from those of routine colonoscopies with fluoroscopy available. Precise detection of loops was observed in the first study in 79 - 100 % of cases in comparison with fluoroscopy. Precise localization of the endoscopic tip improved from 30 % in the first part of the study to 80 % in part 2.
Conclusion: Using magnetic three-dimensional imaging systems, the position of the colonoscope, the detection and observation of loops during straightening, and localization of pathological findings can be accurately achieved. Modification of the prototype led to satisfactory improvement in all parameters tested.

References

  • 1 Bladen J S, Anderson A P, Bell G D. et al . Non-radiological technique for three-dimensional imaging of endoscopes.  Lancet. 1993;  341 719-722
  • 2 Williams C B, Saunders B P, Bell G D. et al . Real-time magnetic three-dimensional imaging of flexible endoscopy.  Gastrointest Endosc Clin N Am. 1997;  7 469-475
  • 3 Shah S G, Saunders B P, Brooker J C, Williams C. Magnetic imaging of colonoscopy: an audit of looping, accuracy and ancillary maneuvers.  Gastrointest Endosc. 2000;  52 1-8
  • 4 Shah S G, Brooker J C, Williams C B. et al . Effect of magnetic endoscope imaging on colonoscopy performance: a randomised controlled trial.  Lancet. 2000;  356 1718-1722
  • 5 Waye J D. Imaging of the colonoscope: magnetic, fluoroscopic, or neither?.  Gastrointest Endosc. 2000;  52 131-133
  • 6 Saunders B P, Bell G D, Williams C B. et al . First clinical results with a real-time electronic imager as an aid to colonoscopy.  Gut. 1995;  36 913-917

P. Frühmorgen, M.D.

Dept. of Medicine, Klinikum Ludwigsburg

Posilipostrasse 4 · 71640 Ludwigsburg · Germany

Fax: + 49-7141-996471

Email: fruehmorgen.peter@t-online.de

    >