Subscribe to RSS
DOI: 10.1055/s-2008-1054877
Gespornter Irishaken zur Implantation von Hinterkammerlinsen vom Typ Kratz, Sinskey und Simcoe
Spurred Iris Hook for Implanting Posterior Chamber Lenses of the Kratz, Sinskey and Simcoe TypesPublication History
Publication Date:
11 February 2008 (online)
Zusammenfassung
Das sanft gespornte Ende des Irisretraktors zwingt die rotierende obere Linsenschlinge hinter die Iris zu gleiten. Der Zwang wird noch verstärkt, wenn das Häkchen um etwa 10-15° nach rechts gedreht wird.
Summary
The iris hook used for posterior chamber implants (e.g., the Kratz, Sinskey and Simcoe types) has a slight spur at its end. The spur catches the upper loop of the rotating lens during the implantation maneuver and forces it to slide behind the iris. If the spurred hook, retracting the pupil upward, is rotated about 10-15° to the right, it guides the loop more safely and more gently into the posterior chamber. The spur prevents the loop from jerking over the iris hook and back into the anterior chamber.