CC BY 4.0 · Surg J (N Y) 2020; 06(01): e42-e46
DOI: 10.1055/s-0040-1702922
Original Article
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Cook–Swartz Doppler Probe Surveillance for Free Flaps—Defining Pros and Cons

1   Department of Plastic, Aesthetic, Reconstructive and Hand Surgery, Agaplesion Diakonieklinikum Rotenburg, Rotenburg (Wümme), Germany
,
Dalius Klimas
1   Department of Plastic, Aesthetic, Reconstructive and Hand Surgery, Agaplesion Diakonieklinikum Rotenburg, Rotenburg (Wümme), Germany
,
Nicco Krezdorn
2   Division of Plastic Surgery, Department of Surgery, Brigham and Women's Hospital, Harvard Medical School, Boston, Massachusetts
,
Dominik Schlarb
3   Department of Plastic and Reconstructive Surgery, Fachklinik Hornheide, Münster, Germany
,
4   Department of Plastic and Reconstructive Surgery, Royal Free Hospital, London, United Kingdom
,
Detlev Hebebrand
1   Department of Plastic, Aesthetic, Reconstructive and Hand Surgery, Agaplesion Diakonieklinikum Rotenburg, Rotenburg (Wümme), Germany
› Institutsangaben
Weitere Informationen

Publikationsverlauf

10. März 2019

16. Januar 2020

Publikationsdatum:
03. März 2020 (online)

Abstract

Introduction The main postoperative complication of free flaps is perfusion compromise. Urgent intervention is critical to increase the chances of flap survival. Invasive flap perfusion monitoring with direct blood flow feedback through the Cook–Swartz Doppler probe could enable earlier detection of perfusion complications.

Materials and Methods Between 2012 and 2016, 35 patients underwent breast reconstruction or defect coverage after trauma with a deep inferior epigastric perforator, anterolateral thigh, transverse musculocutaneous gracilis, gracilis, or latissimus dorsi flap in our department. All flaps were monitored with a Cook–Swartz probe for 10 days postoperatively. The 20 MHz probe was placed around the arterial–venous anastomosis. A flap monitoring protocol was established for standardized surveillance of postoperative perfusion. In the event of probe signal loss, immediate surgical revision was initiated.

Results Signal loss was detected in 8 of the 35 cases. On return to the operating room, six were found to be true positives (relevant disruption of flap perfusion) and two were false positives (due to Doppler probe displacement). There were also two false negatives, resulting in a slowly progressive partial flap loss. Flap perfusion was restored in three of the six cases (50%) identified by the probe. Following surgical intervention, three of the six cases had persistent problems with perfusion, resulting in two total flap losses and one partial flap necrosis leading to an overall 5.7% total flap loss.

Conclusion Postoperative flap perfusion surveillance is a complex matter. Surgical experience is often helpful but not always reliable. The costs, false-positive, and false-negative rates associated with invasive perfusion monitoring with Cook–Swartz probe make it most appropriate for buried flaps.

Level of Evidence This is an original work.

Financial Support and Industry Affiliations

There was no financial support of any kind contributing to this publication. No grants were accepted/used for this publication. None of the authors has any personal or institutional financial interest in drugs, materials, or devices described in this submission.


 
  • References

  • 1 Wax MK. The role of the implantable Doppler probe in free flap surgery. Laryngoscope 2014; 124 (Suppl. 01) S1-S12
  • 2 Whitaker IS, Rozen WM, Chubb D. , et al. Postoperative monitoring of free flaps in autologous breast reconstruction: a multicenter comparison of 398 flaps using clinical monitoring, microdialysis, and the implantable Doppler probe. J Reconstr Microsurg 2010; 26 (06) 409-416
  • 3 Chae MP, Rozen WM, Whitaker IS. , et al. Current evidence for postoperative monitoring of microvascular free flaps: a systematic review. Ann Plast Surg 2015; 74 (05) 621-632
  • 4 de la Torre J, Hedden W, Grant III JH, Gardner PM, Fix RJ, Vásconez LO. Retrospective review of the internal Doppler probe for intra- and postoperative microvascular surveillance. J Reconstr Microsurg 2003; 19 (05) 287-290
  • 5 Rozen WM, Chubb D, Whitaker IS, Acosta R. The efficacy of postoperative monitoring: a single surgeon comparison of clinical monitoring and the implantable Doppler probe in 547 consecutive free flaps. Microsurgery 2010; 30 (02) 105-110
  • 6 Ho MW, Cassidy C, Brown JS, Shaw RJ, Bekiroglu F, Rogers SN. Rationale for the use of the implantable Doppler probe based on 7 years' experience. Br J Oral Maxillofac Surg 2014; 52 (06) 530-534
  • 7 Rozen WM, Enajat M, Whitaker IS, Lindkvist U, Audolfsson T, Acosta R. Postoperative monitoring of lower limb free flaps with the Cook–Swartz implantable Doppler probe: a clinical trial. Microsurgery 2010; 30 (05) 354-360
  • 8 Parker PM, Fischer JC, Shaw WW. Implantable pulsed Doppler cuff for long-term monitoring of free flaps: a preliminary study. Microsurgery 1984; 5 (03) 130-135
  • 9 Swartz WM, Jones NF, Cherup L, Klein A. Direct monitoring of microvascular anastomoses with the 20-MHz ultrasonic Doppler probe: an experimental and clinical study. Plast Reconstr Surg 1988; 81 (02) 149-161
  • 10 Swartz WM, Izquierdo R, Miller MJ. Implantable venous Doppler microvascular monitoring: laboratory investigation and clinical results. Plast Reconstr Surg 1994; 93 (01) 152-163
  • 11 Paydar KZ, Hansen SL, Chang DS, Hoffman WY, Leon P. Implantable venous Doppler monitoring in head and neck free flap reconstruction increases the salvage rate. Plast Reconstr Surg 2010; 125 (04) 1129-1134
  • 12 Guillemaud JP, Seikaly H, Cote D, Allen H, Harris JR. The implantable Cook–Swartz Doppler probe for postoperative monitoring in head and neck free flap reconstruction. Arch Otolaryngol Head Neck Surg 2008; 134 (07) 729-734
  • 13 Lineaweaver WC. The implantable Doppler probe. Plast Reconstr Surg 1988; 82 (06) 1099-1100
  • 14 Lineaweaver W. Techniques of monitoring buried fasciocutaneous free flaps. Plast Reconstr Surg 2009; 124 (05) 1729-1731
  • 15 Rosenberg JJ, Fornage BD, Chevray PM. Monitoring buried free flaps: limitations of the implantable Doppler and use of color duplex sonography as a confirmatory test. Plast Reconstr Surg 2006; 118 (01) 109-113 , discussion 114–115
  • 16 Smit JM, Whitaker IS, Liss AG, Audolfsson T, Kildal M, Acosta R. Post operative monitoring of microvascular breast reconstructions using the implantable Cook–Swartz Doppler system: a study of 145 probes & technical discussion. J Plast Reconstr Aesthet Surg 2009; 62 (10) 1286-1292
  • 17 Schmulder A, Gur E, Zaretski A. Eight-year experience of the Cook–Swartz Doppler in free-flap operations: microsurgical and reexploration results with regard to a wide spectrum of surgeries. Microsurgery 2011; 31 (01) 1-6