Painful wrist arthritis is a debilitating condition for which current treatment options
are limited. The primary goal of treatment for the symptomatic, arthritic wrist is
to achieve a pain-free, functional, and stable joint that is able to bear load. This
is ideally achieved while preserving wrist motion. The causes of wrist arthritis are
varied and include the sequelae of trauma, carpal instability, Keinböck disease, and
inflammatory arthropathy.[1]
[2] The pattern and severity of articular involvement differ according to the etiology
and guide the surgeon in recommending treatment. Ultimately, patients that remain
symptomatic after exhausting conservative options are indicated for one of the several
operative procedures, loosely grouped into ablative surgery, arthrodesis, or arthroplasty.
Ablative procedures are those in which the pain-generating segment of the arthritic
joint is excised. Radial styloidectomy, partial scaphoidectomy, and proximal row carpectomy
(PRC) are the more common ablative procedures performed, with PRC being a recognized
salvage operation for scapholunate advanced collapse (SLAC) and scaphoid nonunion
advanced collapse (SNAC).[3]
[4]
[5]
[6]
[7]
[8]
[9] Ablative procedures, while oftentimes successful in relieving pain, can be complicated
by resultant instability or progressive articular degeneration leading to further
symptoms.[9]
[10]
[11]
[12]
[13]
Total wrist arthrodesis is a reliable operation for pain relief and is favored as
a surgical option in the severely arthritic wrist because of its relative ease of
execution, durability, and predictable long-term results.[14]
[15]
[16]
[17] Pain relief, however, comes at the expense of wrist motion with notable functional
limitations in personal hygiene, applying a forceful grip, picking up small objects,
and working in tight spaces.[18]
[19]
[20]
Partial arthrodesis involves fusing specific segments of the wrist to preserve some
degree of wrist motion while addressing symptomatic arthritic articulations. Limitations
include variable success in fusion rates and the technical difficulties of fusing
small carpal bones. Scaphoid excision and four-corner fusion is the most commonly
performed limited arthrodesis as it preserves a functional arc of motion with satisfactory
grip strength.[21]
[22] Yet, the nonunion rate averages 5%, with reported rates upwards of 25%, and remaining
radiolunate motion is primarily limited to the flexion–extension plane.[22]
[23] Radioscapholunate fusion is another partial arthrodesis for radiocarpal arthritis
and is an attractive surgical option as it preserves critical midcarpal motion and
carpal height.[24] Retaining the midcarpal articulation allows for continued coupled wrist motion known
as the “dart-thrower's” arc of radial-extension to ulnar flexion.[25] Modifications to this procedure, including distal scaphoid and triquetral excision,
have further improved the range of motion and overall outcomes.[26]
[27]
[28] However, problems such as nonunion rates of 3 to 26%, and progressive symptomatic
midcarpal arthritis, are well documented.[29]
[30]
[31]
Wrist arthroplasty is an attractive motion-sparing concept for the treatment of arthritis
and has evolved in multiple iterations over the past 40 years. Swanson first reported
arthroplasty for the wrist in 1973, using a flexible silicone implant.[32] Complications such as implant failure and silicone synovitis fueled the design of
metal-on-polyethylene modular prostheses.[33] Despite advances in design, total wrist prostheses continue to be complicated by
instability and distal component failure.[34]
[35]
[36]
[37]
[38]
[39]
[40]
[41]
[42] Nonetheless, total wrist arthroplasty is a viable option in a carefully selected
patient with inflammatory arthritis of the wrist and low functional demands.
While the mechanics of the wrist remain incompletely understood, growing evidence
has illustrated the importance of the dart-thrower's motion. Studies have demonstrated
that the pattern of carpal bone motion associated with wrist movement along the dart
thrower's path is unique.[43] The mechanical axis of the wrist is not aligned with the anatomic axes of flexion–extension
and radioulnar deviation, but is rather aligned with the dart thrower's axis of radial-extension
to ulnar-flexion.[44] It is hypothesized that all highly functional motions of the upper extremity utilize
the dart thrower's motion. Its presence in many occupational, recreational, and household
activities demonstrates the functional importance of this wrist motion.[45]
The dart-thrower's plane of coupled flexion–extension and radioulnar motion occurs
predominantly at the midcarpal joint, suggesting the importance of this articulation
in prosthetic design.[24]
[43]
[46] Past designs of total wrist arthroplasty have largely replicated the contour and
kinematics of the radiocarpal joint, thus constraining motion to the anatomical directions
and minimizing the important dart thrower's arc. Further, excising portions of the
distal radius relocates the center of rotation (COR) proximal to that of the normal
wrist. There is a substantial clinical evidence that prosthetic failure is associated
with failure to replicate normal joint kinematics in the spine, knee, shoulder, and
the wrist.[47]
[48]
[49]
[50]
[51] A proximal wrist center of rotation would increase the moment on the bone-cement
interface of the distal component and may contribute to the high incidence of distal
component loosening.
The concerning failure rate of current radiocarpal arthroplasty designs, and the improved
understanding of coupled wrist motion, prompted the authors (S.W.W. and J.J.C.) to
design a wrist replacement that would emulate the kinematics of the midcarpal joint.
The KinematX (Extremity Medical, Parsippany, NJ) midcarpal hemiarthroplasty ([Fig. 1]) is a monobloc component that replaces the proximal carpal row, while retaining
the anatomic COR, the normal capsuloligamentous envelope, and preservation of midcarpal
coupled motion. Conceptually, replacement of the arthritic proximal carpal row with
a hemiarthroplasty implant is appealing in that it simultaneously preserves midcarpal
motion and carpal height without the attendant risks of nonunion as seen in partial
arthrodeses or the threat of distal component failure as documented in total wrist
arthroplasty. Furthermore, the technique is potentially easier to execute and leaves
recourse for either total wrist arthroplasty or arthrodesis should revision become
necessary. The purpose of this article is to review the early outcomes of the first
nine cases of midcarpal hemiarthroplasty performed for a variety of etiologies.
Figure 1 The KinematX prosthesis is an anatomical replacement of the proximal row which preserves
the midcarpal articulation and wrist center of rotation.
Indications
Midcarpal hemiarthroplasty is a motion-preserving alternative for the treatment of
painful radiocarpal arthrosis that remains symptomatic and functionally limiting despite
nonsurgical management. Indications include SLAC, SNAC, posttraumatic osteoarthritis,
Keinböck disease, or inflammatory arthritis ([Fig. 2]). Localized radial styloid osteoarthritis can be treated by less invasive means.
The ideal patient for midcarpal hemiarthroplasty is an active patient whose midcarpal
joint is relatively preserved. Patients with ipsilateral shoulder or elbow disease
are also excellent candidates as the preservation of wrist range of motion optimizes
upper extremity function.
Figure 2 SLAC wrist in a 41-year-old active right hand dominant man. Preoperative images above;
postoperative images below at 1 year after surgery. Resumed golfing on a regular basis
at 12 weeks postoperatively. SLAC, scapholunate advanced collapse.
Contraindications/Precautions
Contraindications/Precautions
Successful midcarpal hemiarthroplasty is determined more by individual patient selection
than the underlying diagnosis. This technique is a ligament-sparing method that respects
the complex wrist anatomy to provide coupled wrist motion. As such, the condition
of the surrounding soft tissue envelope and underlying bony architecture is crucial.
The capsuloligamentous sleeve of the wrist must be preserved and patients with severe
soft tissue compromise may be at risk for resultant instability, excluding them as
candidates for hemiarthroplasty. Patients with inflammatory disease and aggressive
active synovitis are likewise at risk for instability or distal carpal row articular
erosion, and may be better served by alternate procedures.
Midcarpal wrist hemiarthroplasty is a prosthetic replacement of the proximal row.
This resurfacing procedure relies on the midcarpal articulation and to what extent
midcarpal arthrosis precludes a good outcome remains unknown. It seems logical that
advanced degenerative disease of the capitolunate joint is a contraindication to midcarpal
hemiarthroplasty but early narrowing may be acceptable. Capitolunate narrowing in
the setting of PRC or distal scaphoid excision has not been shown to compromise results;
and the only published account of wrist hemiarthroplasty reports good pain relief
despite degenerative changes of the capitate head.[6]
[9]
[52]
Durability of the hemiarthroplasty is dependent on prosthetic stability and thus the
distal radius must be capable of supporting the implant. Significant osteopenia, erosions
due to inflammatory arthropathy, or exaggerated epiphyseal/metaphyseal deformity may
adversely affect implant fixation and lead to loosening.
Additional contraindications to hemiarthroplasty include recent or recalcitrant infection,
previous surgical fusion, and lack of active wrist extension. Younger age and active
lifestyle are not considered contraindications, as with total wrist arthroplasty,
since activity restrictions are less stringent given that distal component loosening
is not a concern.
Surgical Technique
The following technique is for implantation of the KinematX midcarpal hemiarthroplasty
(Extremity Medical, Parsippany, NJ) ([Table 1]). Use of a radiographic template is recommended to preoperatively determine appropriate
implant size. The templated implant should reproduce the size of the proximal carpal
row and fit the distal radius on PA and lateral radiographs.
Table 1
KinematX Midcarpal Hemiarthroplasty
Pearls
|
Create a proximally based capsular flap
Leave a wafer of triquetrum attached to capsule to anchor DRC and DIC
Preserve the origins of the palmar and dorsal extrinsic wrist ligaments
Use the power rasp to prepare the surface of the distal radius
Maintain rotational alignment during broaching
|
Pitfalls
|
Failure to place the guidewire down the central axis of the radius
Damage to the capitate head during instrumentation
“Overstuffing” leads to stiffness (small degree of laxity is preferred)
|
DRC, dorsal radiocarpal ligament; DIC, dorsal intercarpal ligament.
Routine skin preparation and upper extremity draping is conducted after the administration
of preoperative IV prophylactic antibiotics. A padded tourniquet is applied to the
upper arm. A dorsal longitudinal incision, 4 to 5 cm in length, is made over the wrist
in line with the third metacarpal. Thick subcutaneous radial and ulnar skin flaps
are elevated directly off the extensor retinaculum to minimize the risk of wound-healing
problems. The extensor retinaculum is incised just radial to Lister's tubercle, reflected
ulnarly, and the fourth dorsal compartment is elevated in continuity from the dorsum
of the radius. At the discretion of the surgeon, the extensor pollicis longus (EPL)
tendon can be freed from its sheath and transposed radially. Additional dissection
of the radius is rarely required, as access to the intramedullary canal is facilitated
by excision of the proximal carpal row. We prefer to preserve the posterior interosseous
nerve as its role in wrist proprioception may prove valuable.[53]
[54]
A wide, proximally based rectangular dorsal capsular flap is created by incising the
capsule beneath the tendons of the second and fifth dorsal compartments. The capsule
is divided distally at the mid-capitate and reflected proximally, leaving a large
capsular sleeve for later reattachment. Using an osteotome, a wafer of the triquetrum
is raised in continuity with the capsule to preserve the attachments of the dorsal
radiotriquetral and dorsal intercarpal ligaments.
With the joint exposed, the proximal row is excised. Removal of the scaphoid, triquetrum,
and lunate can be accomplished in piecemeal fashion or en bloc using joysticks and
sharp curved elevators. It is critical that the surgeon does not damage the palmar
extrinsic ligaments or the cartilage surface of the capitate during removal.
The articular surface of the radius is exposed by wrist flexion and palmar translation
of the distal row. Radial exposure is facilitated with insertion of a thin retractor
under the volar lip. Particular care should be taken to protect the capitate articular
surface at all times. The radius is first prepared using a custom elliptical power
rasp that is contoured to the articular surface of the radius. The intent of the rasp
is to remove the articular cartilage, flatten the interfacet ridge, and shape the
subchondral plate while leaving the origins of the important extrinsic wrist ligaments
fully intact dorsally and volarly. The tool is best used in a radial to ulnar and
dorsal to volar sweeping motion. The subchondral bone provides support for the implant
and should not be removed.
A 2.5 mm primary guidewire is inserted down the medullary canal, beginning ~5 mm volar
to the dorsal lip of the distal radius, or between the middle and dorsal thirds of
the radial articular surface height. The ulnar border of Lister's tubercle is an anatomic
landmark to guide proper placement in the coronal plane. Central positioning of the
guidewire is verified under fluoroscopy.
Canal preparation begins with creating a window through the subchondral plate for
broaching. A box chisel corresponding to the templated implant size is placed over
the guidewire and impacted with the mallet until seated. The portion of the subchondral
bone designated by the box chisel outline may be removed using small osteotomes or
a high speed burr. The starter broach is impacted over the guidewire until flush with
the subchondral bone. The canal is sequentially broached until the templated size
is reached. Maintaining appropriate alignment during broaching is critical. Malrotation
may cause the ulnar and radial borders of the implant to lift off of the supporting
subchondral bone. The guidewire is removed once broaching is complete.
The trial component is impacted into position and the carpus is gently reduced onto
the bearing surface of the implant, taking care to avoid injury to the capitate articular
surface. Implant size, joint range of motion, and stability are assessed. In general,
a degree of laxity is preferred over tension in the dorsal and volar ligaments, and
will lead to more rapid return of motion. The final implant is gently impacted until
it is fully seated on the contoured subchondral plate. Pressurized cementing is an
option if compromised bone prohibits press fit technique, although we have not found
this to be necessary. The carpus is reduced and the capsule repaired anatomically
with a single running suture. The extensor retinaculum is repaired and the EPL transposed
at the surgeon's discretion. The wound is closed in layers with or without suction
drainage, and a volar plaster wrist splint applied in neutral position.
The postoperative protocol for midcarpal hemiarthroplasty is relatively straightforward.
Active digital, shoulder and elbow range of motion is initiated on postoperative day
1. The patients return to the office 10 to 14 days after surgery for suture removal.
Immobilization is discontinued at that time and patients are enrolled in a supervised
program of progressive wrist range of motion (flexion, extension, radial and ulnar
deviation, coupled motion including dart-throwing, pronation, supination, and circumduction).
Strengthening begins 4 to 6 weeks from the time of surgery and full activity is permitted
at 8 weeks. There are no permanent activity restrictions once strength is returned.
Potential Complications
Wrist hemiarthroplasty assumes the following general risks of implant arthroplasty:
wound healing problems, infection, hematoma, aseptic loosening, joint stiffness, nerve
or tendon injury, and intraoperative fracture. Potential complications specific to
the wrist include extensor adhesions, wrist instability, carpal impingement, and development
of symptomatic midcarpal arthrosis. The extent to which these potential complications
will play a role in midcarpal hemiarthroplasty is yet to be determined.
Outcomes
Restoration of wrist function, while an obvious goal of wrist reconstructive surgery,
has been suboptimal in the treatment of painful arthritis of the wrist. Hemiarthroplasty
is a novel approach and provides patients with a motion-preserving alternative to
procedures that focus on pain relief at the expense of wrist biomechanics and natural
motion. This technique restores anatomy by resurfacing the proximal row to preserve
the midcarpal joint and the anatomic center of wrist rotation. Carpal kinematics research
has demonstrated that this crucial articulation is key to the highly functional dart-thrower's
motion of radial-extension to ulnar-flexion.[43]
[44]
[45]
[46]
[55]
Wrist hemiarthroplasty is a new technique and current literature is accordingly limited.
The only published account of the procedure details the experience of a single surgeon
that implanted the proximal component of a Universal 2 Total Wrist (Integra Life Sciences,
Plainsboro, NJ) in two young, active patients who were not candidates for total wrist
arthroplasty. At 1 and 2 years after surgery, each patient remained pain free, resumed
unlimited activity, and maintained range of motion comparable to preoperative levels
(flexion arcs of 80 and 69 degrees, respectively). However, there was evidence of
hamate erosion in the second case, presumably due to prosthetic impingement, which
may compromise long-term success.[52] Using a cadaveric model, Adams et al showed retention of static capitate alignment
after PRC and hemiarthroplasty. Furthermore, their series of 12 patients had no evidence
of dislocation or prosthetic complications at early follow-up (“Distal radius arthroplasty
with PRC using the Universal 2 implant.” International Wrist Investigator's Workshop. 2010).
The KinematX midcarpal hemiarthroplasty was designed to emulate the anatomic contour
of the native midcarpal joint, and the corresponding surgical technique developed
to preserve the anatomic capsuloligamentous and neural anatomy. The prosthesis preserves
radial length and maintains the anatomic COR to provide coupled wrist motion. Replacement
of the diseased proximal row provides a painless, stable midcarpal platform for activities
requiring combined force and precision.
Over the past 13 months, we have implanted the KinematX hemiarthroplasty in nine patients.
Prospective outcome data are currently being collected with Institutional Review Board
approval and includes both validated subjective questionnaires and objective measurements.
The average age at the time of midcarpal hemiarthroplasty was 43.6 years (range: 23
to 74 years.) Three patients were male and six were female. The dominant hand was
involved in six patients. Seven of the nine patients treated were working at the time
of surgery, none of which were manual laborers. The average total surgery time was
49 minutes (range: 45 to 60 minutes).
Before surgery, all patients were suffering from chronic, painful degenerative wrist
arthritis that limited use of the hand for daily and recreational activities. Posteroanterior
and lateral wrist radiographs confirmed the diagnosis in all patients and advanced
imaging was not required. The indications for surgery were as follows: SLAC stage
2 (one), SLAC stage 3 (one), SNAC stage 3 (one), posttraumatic osteoarthritis(three),
inflammatory arthritis (two; rheumatoid arthritis and psoriatic arthritis), and Keinböck's
stage 4 (one). Two patients had preoperative radiographic evidence of early joint
space narrowing of the capitolunate articulation. Prior surgery for wrist pain had
been performed in two patients (radial styloidectomy for SLAC I, distal radioulnar
joint arthroplasty for posttraumatic osteoarthritis). Preoperative range of motion,
grip strength, DASH, and Mayo wrist scores are shown in [Table 2].
Table 2
Pre- and Postoperative Data[a]
|
Preoperative
|
Postoperative
|
Significance (p)
|
Mayo wrist score
|
31.9 (10–60)
|
58.8 (30–80)
|
0.006
|
DASH score
|
47.8 (22.7–70.5)
|
28.7 (0–68.2)
|
0.028
|
FE arc, degrees
|
64.6 (40–125)
|
79.3 (30–130)
|
0.362
|
RD-UD arc, degrees
|
16.9 (5–50)
|
22.9 (5–37)
|
0.262
|
Grip, kg
|
16.1 (6–35)
|
18.9 (6–38)
|
0.496
|
Grip, % of opposite side
|
56.3 (30–77.8)
|
61.7 (31–91)
|
0.501
|
Note: Includes all patients who underwent hemiarthroplasty (posttraumatic and inflammatory
etiologies). Data reported as mean value (range).
a All patients.
DASH, disabilities of the arm, shoulder, and hand; FE, flexion extension; RD-UD, radial
deviation-ulnar deviation; kg, kilograms.
All nine patients were interviewed and examined at regular intervals during their
postoperative course. The average length of follow-up was 30.9 weeks (range: 16 to
56 weeks). The mean Mayo wrist score increased from 31.9 (range: 10 to 60) preoperatively
to 58.8 (range: 30 to 80; p < 0.05) and the mean disabilities of the arm, shoulder, and hand (DASH) score improved
significantly from 47.8 (range: 22.7 to 70.5) preoperatively to 28.7 (range: 0 to
68.2) at latest follow-up (p < 0.05). The index patient reported a DASH score of 0 at 1 year postoperatively and
has resumed golfing regularly. Five of the seven patients that were working before
surgery had returned to their regular occupation.
Although there was a trend toward increased motion after surgery, there was no significantly
detectable difference in pre- and postoperative range of motion or grip strength measurements
([Table 2]). Mean flexion–extension arc was 79 degrees (range: 30 to 130 degrees) and 52.8%
of the opposite hand. Radioulnar deviation averaged 22.9 degrees (range: 5 to 37 degrees)
and 33.6%. Mean grip strength was 18.9 kg (range: 6 to 38 kg), which corresponded
to 61.7% of the nonsurgical side.
To date, there was no detectable change in subsequent postoperative radiographs. Neither
progressive capitolunate degeneration nor signs of implant loosening were observed.
The only complication identified in our series was postoperative wrist stiffness requiring
manipulation under anesthesia in two patients (patient 4 with SLAC wrist and patient
8 with psoriatic arthritis). Their postoperative flexion–extension arcs were restricted
to 0 and 5 degrees, respectively, which improved to 50 and 30 degrees after manipulation.
Open contracture release was not required. Of note, each of these patients was severely
limited in their preoperative motion (45 and 40 degrees), which possibly contributed
to their resultant stiffness.
Of specific interest, the two patients in the series with inflammatory arthropathy
performed poorly across all points and time. Their postoperative Mayo wrist[30]
[35] and DASH (65.9 and 68.2) scores, and grip strengths (8 kg in both) indicate persistent
functional limitation not observed in the posttraumatic patients. This prompted us
to closely re-examine our data, with the inflammatory patients eliminated from the
analysis, and discovered that postoperative increase in motion reached statistical
significance. We also observed that the mean postoperative Mayo wrist and DASH scores
improved further ([Table 3]). Perhaps this is due to unrecognized and persistent inflammatory activity not addressed
by the proximal row replacement. While it is not possible to draw a firm conclusion
based on the early results of the two patients, we may learn that this procedure is
best suited for the patients with posttraumatic degenerative disease.
Table 3
Data for Posttraumatic Patients (Inflammatory Arthritis Excluded)
|
Preoperative
|
Postoperative
|
Significance
|
Mayo wrist score
|
35
|
67.5
|
0.006
|
DASH score
|
43.2
|
15.9
|
0.006
|
FE arc, degrees
|
58.7
|
90.8
|
0.039
|
RD-UD arc, degrees
|
13.3
|
24.7
|
0.035
|
Grip, kg
|
17.8
|
22.5
|
0.217
|
Note: Data reported as mean value.
DASH, disabilities of the arm, shoulder, and hand; FE, flexion extension; RD-UD, radial
deviation-ulnar deviation; kg, kilograms.
The limitations of our study must be acknowledged when reviewing our results. Our
data are based on single cohort of patients, which presents potential inherent bias.
The brief follow-up period precludes any determination of implant durability.
Hemiarthroplasty of the wrist is an exciting and novel technique which is still in
its infancy. The procedure is simple and maintains coupled wrist motion through preservation
of the midcarpal articulation. The preliminary data in our series support other reports
that this technique, in the early postoperative period, appears safe, and has demonstrated
encouraging early results. We acknowledge that longitudinal tracking of these patients
is required before conclusions can be drawn as to its durability, reliability, and
overall benefit. Further study is currently underway to determine whether midcarpal
hemiarthroplasty represents an advancement of motion-preserving technology for the
treatment of painful radiocarpal arthritis.